background image

 

D-0123721-A – 2019/05 

 

Page 

1

 of 

3

 

 

Quick Guide 

Listen to coupler / Listen to probe microphone 

The purpose of this document is to provide a quick guide for the Listen to Coupler/Listen to Probe Microphone 

feature found in the REM440 Real-Ear Measurement module in the Affinity

 

Suite. This Quick Guide will focus on 

how to use this function.  

Introduction 

It can sometimes be difficult to understand feedback from your patient regarding how their hearing aid sounds. 

The Affinity Compact has a handy feature which allows you to listen directly to the probe tube microphone or 

coupler microphone to hear exactly what the hearing aid is doing.  

 

Preparation 

1.  Enter your REM module as you would normally

.

 Ensure you have a monitor headset plugged into your 

Affinity Compact, the socket which this is connected to is shown below.  

 

2.  Connect 

probe tubes

 to the REM reference microphones. 

3.  Click on the 

Tube Calibration

 button  

  and follow the instructions on the screen, this process ensures 

that the probe tube is made acoustically invisible – enabling an effective display of the actual gain achieved 

at the tympanic membrane by the hearing aid.  