Hardware Manual: Part Number 10554 Rev C., iQue Maintenance: Fluidics Maintenance
Removing the Optics Cover
Prior to performing the first two steps of the fluidics maintenance procedure, the optics cover must be removed from
the detector. It is not necessary to shut down the detector before doing this as the optics cover contains an interlock
that automatically keeps the system from running when the cover is off.
1.
Open the lid of the iQue to expose the detector area. Locate the optics cover.
Optics Cover
2.
Using the 3mm hex key that came with the system, loosen the two captive screws that secure the optics
cover.
3.
Gently lift the optics cover off the detector area.
Содержание iQue SCREENER
Страница 1: ...iQue SCREENER hardware manual October 2014...
Страница 2: ......
Страница 11: ...5 iQue Hardware Manual Part Number 10554 Rev C Introduction Declaration of Conformity...
Страница 12: ...6 THIS PAGE INTENTIONALLY LEFT BLANK iQue Hardware Manual Part Number 10554 Rev C Introduction...
Страница 16: ...10 THIS PAGE INTENTIONALLY LEFT BLANK iQue Hardware Manual Part Number 10554 Rev C Using This Manual...
Страница 22: ...16 THIS PAGE INTENTIONALLY LEFT BLANK iQue Hardware Manual Part Number 10554 Rev C Safety...
Страница 34: ...28 THIS PAGE INTENTIONALLY LEFT BLANK iQue Hardware Manual Part Number 10554 Rev C iQue Operations Cover Page...
Страница 46: ...40 THIS PAGE INTENTIONALLY LEFT BLANK iQue Hardware Manual Part Number 10554 Rev C iQue Operations Daily QC...
Страница 50: ...44 THIS PAGE INTENTIONALLY LEFT BLANK iQue Hardware Manual Part Number 10554 Rev C iQue Maintenance Cover Page...
Страница 64: ...iQue Hardware Manual iQue Maintenance 58 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 90: ...84 THIS PAGE INTENTIONALLY LEFT BLANK iQue Hardware Manual Part Number 10554 Rev C Special Topics Cover Page...
Страница 110: ......