55
5
Or press
Back
to return to the previous screen.
n
Check Pod Status
To check Pod status, if the PDM is turned
Off
, press and hold the
Power
button to display the
ID
screen, then press
Confirm
to
display the
Status
screen. If the PDM is
On
, press
Back
until you
reach the
Status
screen. (If you press and hold the
Power
but-
ton, you will turn off the PDM instead.) Or, from the
Home
screen, press
Status
.
The PDM automatically checks the status of the Pod. The PDM
then displays the current insulin reservoir volume, PDM battery
level, last blood glucose value, last bolus dose, active basal pro-
gram, and any alarm conditions. See Chapter 6, Using the
Personal Diabetes Manager, for details of what you see on the
Status
screen.
If the Pod goes into hazard alarm condition and the PDM is
unable to communicate with it to turn off the alarm, you can turn
it off manually on the Pod. See the end of Chapter 10, Alerts and
Alarms, for instructions.
NEVER inject insulin (or anything else) into the
fill port while the Pod is on your body. Doing so may
result in unintended or interrupted insulin delivery.
Figure 5-19
Figure 5-23
Once a Pod is activated and communicates with the
PDM, it can only receive commands from that PDM,
not from any other.