Страница 1: ...www InstroTek com OPERATING MANUAL innovators in instrumentation technology Rapid Asphalt Drying System COREDRY...
Страница 2: ...2018 InstroTek Inc CoreDry Operation Manual Version 8...
Страница 3: ...use PRECAUTIONS WHEN USING THE COREDRY 1 There is an air heater located inside the CoreDry to maintain the sample at room temperature The heat is on during flow cycle THE BOTTOM OF THE DRYING CHAMBER...
Страница 4: ...you to do this Only use recommended vacuum oil InstroTek part number 419 0005 6 Change your Tank Filters every 1 to 2 months depending on usage InstroTek part number 977 1004 package of 12 filters 7 C...
Страница 5: ...InstroTek 1 1 Table of Contents 1 INTRODUCTION 2 2 GETTING STARTED 5 3 MENU FUNCTIONS 10 4 AGGREGATE TESTING 12 5 EMULSION TESTING 15 6 MAINTENANCE 20 7 TROUBLESHOOTING 26 8 INDEX 28 9 WARRANTY 30...
Страница 6: ...user with samples that can be tested in a few minutes Now you can use the dry weight from the CoreDry to measure your core density in minutes or get a quick measure of your stockpile aggregate moistu...
Страница 7: ...CoreDry Components CoreDry Front View CoreDry Back View Cold Trap with Lid Sample Chamber with Lid On Off Switch Oil Fill Plug Oil Level Site Glass Glass Drain Valve Power Modules Serial Number Plate...
Страница 8: ...InstroTek 4 Flow Plate Wire Mesh Sample Flow Plate and Sample Holder Installed in Sample chamber Tank Filters...
Страница 9: ...e the Sample Chamber lid which is located next to the CoreDry unit inside the box The Flow Plate Mesh holder is in the bottom of the large tank along with extra tank filters the Cold Trap lid a bottle...
Страница 10: ...X X System Pressurizing XX CoreDry Version X X X Dry Cold Trap T XX CoreDry Version X X X Cold Trap Cooling XX System Ready P1 Start to Begin Auto Mode Program X Cycle XX T XX Pressure XXX mmHg Auto...
Страница 11: ...e information Note When you receive the CoreDry or after transport to other locations you might have to run the CoreDry several cycles to achieve 6 mmHg Moisture from humidity might build up within th...
Страница 12: ...phalt sample to remove as much of the free water as possible Place the sample on its side then on the wire mesh sample support as shown below place the lid on sample chamber and press start When sampl...
Страница 13: ...d trap lid and the divider plate and wipe out the moisture Be sure to always replace the divider before drying the next sample Caution Use a lint free cloth to remove moisture from the cold trap and t...
Страница 14: ...yed in hours 3 Run Hours Displays the amount of time the CoreDry unit has operated Time is displayed in hours 4 Max Cycles This menu function shows the maximum number of cycles the CoreDry will run un...
Страница 15: ...oil changes back to 80 hours After each oil change it is important to go into this menu function and reset the timer 9 Tank Filter Resets the timer for the tank filters After the tank filters have bee...
Страница 16: ...ggregate quantities to use in the fixture Record weight 5 When you determine the sample weight re tare the scale 6 Replace the lid and tighten the fasteners on the aggregate fixture 7 Weight the entir...
Страница 17: ...e inside the sample chamber Close the lid and press Enter The unit will continue until the sample is completely dry 12 Determine the moisture content by the following equation Moisture Content Weight...
Страница 18: ...InstroTek 14 Aggregate Sample Fixture Aggregate Fixture with one fastener and lid removed...
Страница 19: ...nutes using the CoreDryTM It enables laboratories to recover the residual asphalt of emulsion to determine the percent residual asphalt without changing the binder properties The residual binder can b...
Страница 20: ...reDry control board necessary for older models Preparing a Sample 1 Prepare a space with a protective sheet to capture any spills of the emulsion 2 Weigh a clean silicone sheet and metal pan to the ne...
Страница 21: ...ion The applicator should slide across the surface of the silicon sheet 5 Once the emulsion is spread place the silicone sheet back in the metal pan Be careful not to spill the emulsion 6 Weigh the pa...
Страница 22: ...tion process for the remaining 4 pans 10 Turn on the CoreDry and select Program 5 for drying the emulsion a Enter Menu Select option 1 Program Press Enter b Press the Up Down arrows until 5 is shown c...
Страница 23: ...sides together to form a ball of asphalt binder NOTE If the residual binder is to be tested use latex or neoprene gloves to remove the binder to prevent contamination of the sample 2 Wash the silicone...
Страница 24: ...ared and after you change the filters go into the Menu option 9 to reset the timer for the next change Change Oil Every 80 hours of normal use this will be indicated on the LCD readout and can be chec...
Страница 25: ...dispose of the oil in a responsible manner Always discard the oil at certified disposal facilities Caution Carefully remove any spilled oil from the floor to avoid accidents Only use high quality syn...
Страница 26: ...edges and 1 on each side in the middle near the bottom 4 Prior to lifting top cover place a piece of foam or other support in front of CoreDry 5 Lift top cover and rotate towards the front to expose t...
Страница 27: ...and top cover 14 Replace all exterior screws 15 Go into Menu option 10 and reset the timer for the Exhaust Filter Changing Pump Filter Welch Pump Replacement part number 977 1002 1 Turn the ON OFF swi...
Страница 28: ...up CoreDry by running the unit empty with both lids on for three separate tests 2 Dry Cold Trap to make sure there is no moisture inside the Cold Trap 3 Place Vacuum Gauge inside sample chamber and p...
Страница 29: ...25 7 If the two readings are not the same take note of the difference between the Vacuum Gauge and the displayed pressure and contact InstroTek at 919 875 8371 for instruction on how to adjust pressu...
Страница 30: ...sed completely Run Self Test to make sure everything is working within normal parameters Open CoreDry and make sure no hoses are pinched or loose Contact InstroTek for troubleshooting help 919 875 837...
Страница 31: ...on each side in the middle near the base Rotate top cover forward and slowly remove one of the hoses from sample tank WARNING DOING SO WILL RELEASE VACUUM IN TANK AND WILL CAUSE THE LIDS TO BECOME FR...
Страница 32: ...egate Type 13 Auto Mode 6 B Busch Pump 21 C Changing Oil 15 Components 3 D Daily Test 7 Drain Valve 26 Dry Plate 4 Drying Samples 8 E Emulsion Cleaning 19 Emulsion Testing 15 F Flow Plate 5 M Maintena...
Страница 33: ...reparing a Sample 16 Prep Flow Chart 6 Pump Filter 21 23 Q Quick Start 5 S Self Test 10 11 26 Setup 5 T Tank Filters 4 5 11 20 Tools Needed 22 Troubleshooting 26 V Verifying Vacuum Pressure 24 Views 3...
Страница 34: ...ipping charges duties or taxes in the transport to and from the factory or authorized service center Always insure the shipment if a return for repair is needed InstroTek does not take responsibility...
Страница 35: ...1 Contact Information Innovators in Instrumentation Technology www InstroTek com 919 875 8371 sales InstroTek com Research Triangle Park NC Austin TX Bensalem PA Grand Rapids MI Denver CO Las Vegas NV...