3119-600 Series Temperature Controlled Chambers

Operating Instructions

M55-16547-EN Revision 

E

The difference is measurable

 ®

Содержание 3119-600 Series

Страница 1: ...3119 600 Series Temperature Controlled Chambers Operating Instructions M55 16547 EN Revision E The difference is measurable...

Страница 2: ...ormation contained therein may be acquired only by written permission signed by a duly authorized officer of ITW Trademarks Instron is a registered trademark of Illinois Tool Works Inc ITW Other names...

Страница 3: ...afety 11 General Safety Precautions 11 Warning Signs Used 12 Electrical 13 Thermal Contact 14 Specimens 14 Cryogenic Gases Asphyxiation and Toxicity 15 Glass Fibre Insulation 15 High Pressure Gases 16...

Страница 4: ...erature 31 Shutting Down the Chamber 33 Preventive Maintenance 34 Changing the Eurotherm 3208 Controller Menu 35 Changing the Alarm Settings 36 Auxiliary Alarm 36 Resetting a Tripped Alarm 37 Changing...

Страница 5: ...et and DC Output Calibration 55 Resetting the Chamber to Factory Defaults 55 Calibrating an External Device using the Controller s DC Output 55 DC Output Configuration 56 Appendix B Hints and Tips for...

Страница 6: ...Preliminary Pages 6 M55 16547 EN...

Страница 7: ...th removable wedges to allow easy removal of the chamber from the test area without dismantling the load string Different height chambers are available these are selected to suit the required applicat...

Страница 8: ...ion certification and maintenance of the eyebolts as lifting equipment become the customer s responsibility Refer to relevant local legislation Figure 1 3119 600 Chamber Components Heated Window Door...

Страница 9: ...performance of the controller or cause damage to the chamber The procedures detailed in this manual have been validated by Instron You are advised to use only these specific procedures and in the man...

Страница 10: ...is stamped onto a serial number plate that is attached to the rear of chamber above the rear input panel Control Panel and Rear Input Panel Labelling See page 20 for control panel labelling and page...

Страница 11: ...7 General Safety Precautions Warning Hazard Chambers have inherent critical risks you are strongly advised to carry out your own safety risk assessment taking into account all the information provided...

Страница 12: ...carry out your own product safety risk assessment and use this to determine suitable personal protective equipment Warning Signs Used When a symbol is enclosed in a triangle it becomes a warning sign...

Страница 13: ...vers are removed Refit the covers as soon as possible Electrical Disconnection Hazard After disconnection from the power supply leave adequate time for any residual charge to dissipate before touching...

Страница 14: ...ndling of cryogenic gases from your supplier High Low Temperature Hazard Wear adequate personal protective equipment when using equipment at extremes of temperature This hazard applies to the physical...

Страница 15: ...upplier s Material Safety Data Sheet for limits and monitoring information Toxicity Hazard Carbon dioxide is toxic In certain concentrations carbon dioxide can be dangerous Refer to your supplier s Ma...

Страница 16: ...local Instron representative if you suspect a fault with the chamber There are exposed rotating parts of the fan in the rear electrical compartment of the chamber In normal operation access to the re...

Страница 17: ...s only Microtherm G block is contained between the 3119 608 chamber s outer panels and inner case Belgium MICROTHERM N V INDUSTRIEPARK NOORD 1 B 9100 SINT NIKLAAS Phone 32 0 3 760 19 80 Fax 32 0 3 760...

Страница 18: ...tment Route de Lauterbourg BP 148 67163 WISSEMBOURG Cedex France Phone 33 0 3 88 54 95 50 Fax 33 0 3 88 54 29 20 Blanket MSDS No 105 Paper MSDS No 357 To confirm this is the most up to date MSDS for t...

Страница 19: ...2 Removable Wedges 24 Chamber Door Power Indicator The white LED power indicator is located at the top right hand side of the door It illuminates when the circuit breaker is switched on see Rear Input...

Страница 20: ...cribed below Figure 2 Control Panel Eurotherm 3208 Temperature Controller Page Button Heating Button Indicator above Door Open Indicator Up Button Cooling Button Indicator above Scroll Button Down But...

Страница 21: ...operating parameters See the procedures in the Operating the Chamber chapter SPX Indicator Not used by any procedures detailed in the chamber manuals If illuminated contact your local Instron represen...

Страница 22: ...tes GREEN when the light is ON Door open indicator When the door is OPEN the indicator illuminates RED The fan and all heating and cooling operations are halted when the door is opened They resume whe...

Страница 23: ...arm switch input such as a water cooled grips flow switch An alarm condition will be detected when the switch contacts are open When not required use shorting plug supplied 3119 607 3119 608 and 3119...

Страница 24: ...ore should always be handled with care Removing the Wedges 1 Open the chamber door 2 Grip the wedge using the hand recess and pull forwards 3 Repeat the above procedure for the remaining wedge Replaci...

Страница 25: ...the Ramp Rate 38 Using the Timer Programmer 39 Connecting the Chamber to a Computer 43 Analog Output 45 Fault Finding 45 Caution Other than the procedures in this chapter do not adjust the configurat...

Страница 26: ...t system use Always clip or tie cables to avoid a trip hazard You are familiar with the information in Appendix B Hints and Tips for Using Chambers If cooling is required ensure that The room is suita...

Страница 27: ...ounterclockwise as far as it will go and is vertical Crush Hazard Moving components can cause injury After the chamber push pullrods and grips fixtures etc have been installed it is important to preve...

Страница 28: ...icator extinguishes Enabling Low Speed Fan Mode There are two modes of fan operation full speed and low speed In normal use the fan is set to full speed mode by pressing the fan button for less than 1...

Страница 29: ...o in the following procedures as the home page Checking Settings If you wish to change the units of temperature displayed see Changing the Temperature Units on page 37 If you are the first user the ch...

Страница 30: ...GREEN when the fan is running If it is not illuminated press the fan button to start the fan 4 To begin thermal control either a Press the heater button or b If the setpoint is below ambient temperatu...

Страница 31: ...oling To stop heating or cooling deselect the appropriate button Switching the fan off will stop heating and cooling automatically Returning the Chamber to Ambient Temperature The best way to complete...

Страница 32: ...ter Operating the Chamber 32 M55 16547 EN Caution If you do not allow the chamber to return to ambient temperature after testing below 0 C ice may form on the inside of the chamber when you open the d...

Страница 33: ...mber 1 If you are using high pressure cryogenic gas Warning a Shut the high pressure cryogenic gas cylinder valve b Set the chamber to a cooling cycle The solenoid valve opens and the pressure in the...

Страница 34: ...ational or international standards Contact your local Instron representative If the chamber is used routinely below 0 C then the thermocouple characteristics can change more rapidly In this case it is...

Страница 35: ...section are in the order they appear in the menu There are additional menu choices when tm cfg has been configured to prog See Using the Timer Programmer on page 39 for details Figure 5 Eurotherm 320...

Страница 36: ...the down button as appropriate until the figure displayed is the alarm value you require 3 Wait a few seconds The lower display flashes once to show the value is accepted 4 Press the page button to re...

Страница 37: ...the control panel alarm indicator You can then start heating or cooling again Changing the Temperature Units Temperature units are displayed at the top right hand of the controller display To change t...

Страница 38: ...ng and specimen can have an affect the speed of temperature changes The default ramp rate setting OFF applies full power to the heating element or fully opens the cooling solenoid where optional cooli...

Страница 39: ...until OFF is displayed 8 Follow steps 3 and 4 The RUN indicator will no longer be displayed when the setpoint is changed The setpoint ramp rate is now disabled and the chamber will heat and cool at i...

Страница 40: ...press the scroll button a number of times until TM CFG is shown in the lower display 2 Press either the up button or the down button until nonE appears in the controller upper display 3 Wait a few se...

Страница 41: ...1 scrolls across the screen 2 Press the scroll button Tsp 1 appears in the lower display If you wait a few seconds the text Target setpoint1 scrolls across the screen Figure 7 Example of Eight Segment...

Страница 42: ...2 From the home page press the scroll button a number of times until T STAT is shown in the lower display 3 Press either the up button or the down button to select either a run The program starts and...

Страница 43: ...driver Before Installation 1 Do not connect the chamber USB to the computer 2 Log in as an administrator Installation 3 Insert the CD 4 Run setup exe 5 Connect the USB cable to the chamber and a spare...

Страница 44: ...tem and Maintenance Device Manager Windows 10 Find and open the Device Manager b If you are prompted for an administrator password or confirmation type the password or provide confirmation c Expand Po...

Страница 45: ...o Table 5 to identify the possible cause s Any repairs that require the dismantling of the chamber should only be carried out by an Instron service engineer Should a fault be discovered with your envi...

Страница 46: ...ck that the mains O I switch is set to I If the problem persists contact your local Instron representative The push buttons do not illuminate and the push buttons have no effect but the display illumi...

Страница 47: ...If the problem persists contact your local Instron representative Lights do not illuminate Check bulbs and replace if necessary If the problem persists contact your local Instron representative Alarm...

Страница 48: ...Chapter Operating the Chamber 48 M55 16547 EN...

Страница 49: ...x details how to change chamber control parameters If such changes lead to problems operating the chamber you should restore the factory settings The product warranty does not cover user changes to th...

Страница 50: ...peration Level 1 Setup Level 2 1 Autotune To improve control at specific temperatures 2 Not Available Read Write 2 Control Parameters To allow manual adjustment of control behaviour 8 Not Available Re...

Страница 51: ...now in level 2 1 Press and hold the page button After a few seconds the upper display shows LEv2 and the lower display shows GOTO 2 Press the down button to select LEv1 The controller returns to level...

Страница 52: ...TM CFG AUTO TUNE ENABLE TO ADJUST ALARM 2 SETPOINT TO ADJUST DISPLAY UNITS TO VIEW OPTIONS SETPOINT RATE LIMIT TO ADJUST TIMER CONFIGURATION TO VIEW OPTIONS CONTROL PARAMETERS PROPORTIONAL BAND etc RE...

Страница 53: ...ctory settings see page 55 To autotune 1 Ensure that your controller is set to level 2 see page 51 the chamber is close to ambient temperature and the fan is disabled 2 From the Home Page press the up...

Страница 54: ...or the auto tuning function see page 53 do not give you the performance you require visit www eurotherm com downloads documents user guides and download the 3200 Engineering Manual for more details a...

Страница 55: ...s the scroll button until rec no appears in the lower display The upper display will show nonE If you wait a few seconds the text current recipe number scrolls across the screen 3 Press the up or down...

Страница 56: ...low taking the lowest value 10 o C 18 o F as 0 V to allow for overshoot Your device is now calibrated 9 Run the reset recipe See Resetting the Chamber to Factory Defaults on page 55 10 Press the scrol...

Страница 57: ...energy consumption The window heaters are disabled by switching off cooling Also note that window heaters themselves can result in the inside temperature of the chamber rising well above ambient temp...

Страница 58: ...ur Equipment Make sure you use the recommended anti seize compounds and lubricants on your grips to help aid specimen removal and minimise corrosion due to condensation Refer to the documentation supp...

Страница 59: ...mber the specimen temperature will not follow immediately Consider using an independent temperature sensor attached to or very near the specimen for a more accurate measurement of specimen temperature...

Страница 60: ...um the thermocouple temperature is usually a close representation of the specimen temperature If you are especially concerned about the specimen temperature then a suitable measurement thermocouple sh...

Страница 61: ...let where available Using this set up may require the control parameters to be changed see Autotune on page 53 This set up may prevent the chamber from reaching its lowest or highest rated temperature...

Страница 62: ...Appendix Hints and Tips for Using Chambers 62 M55 16547 EN...

Страница 63: ...hamber door 27 connecting the chamber to a computer 43 connections rear panel 22 control parameters 54 controller menu 35 controller menu level 2 51 controls and indicators 20 cooling stopping 31 to s...

Страница 64: ...ng heating or cooling 31 switching on 28 P power indicator 19 preventive maintenance 34 product labelling 10 program how to run 42 programmer 39 R ramp rate changing 38 rear panel connections 22 recip...

Страница 65: ...ilation air circulation 60 W warning asphyxiation 15 cryogenic gases 15 electrical 13 electrical disconnection 13 glass fibre insulation 15 high pressure gases 16 moving parts 16 rotating machinery 16...

Страница 66: ...W 66 M55 16547 EN...

Страница 67: ......

Страница 68: ...Product Support www instron com...