![Instron 1000KPX-J3 Скачать руководство пользователя страница 42](http://html1.mh-extra.com/html/instron/1000kpx-j3/1000kpx-j3_system-concepts-manual_2070124042.webp)
Chapter 3: Additional System Details
42
M47-17036-EN
Hydraulic system
Introduction
Warning
The hydraulic system includes any component that controls or routes hydraulic fluid (oil) through the
testing system. For specifications of the hydraulic oil, refer to maintenance information provided in the
System Operating Instructions (supplied separately).
Hydraulic power supply
The function of the HPS is to act as the power supply for the frame. This includes powering the hydraulic
cylinder and hydraulic lift and lock cylinders (for frames with a J3-style crosshead). The HPS is a variable
pressure unit; it creates hydraulic pressure as required depending on system demand.
The HPS is enclosed in a console that has a desktop surface for placement of the computer system and
59 Series control unit. The components of the HPS can vary depending on the configuration of the
system. In general however, major components of the HPS are: a reservoir; a pump; an electric motor; a
heat exchanger; one or more manifold assemblies; and the hydraulic fluid (oil).
identifies
typical HPS components. An electrical panel mounted inside the console houses the motor starters, with
overloads, for the electric motor.
When the HPS is on, the hydraulic cylinder should be positioned so that the piston is neither fully
retracted nor fully extended. When the hydraulic cylinder is in either of these conditions, it is difficult for
the 59 Series controls to maintain the position and the servo loop control could build up enough error to
shut down the HPS.
Reservoir
Caution
DO NOT start the HPS without the screen trapper or air breather assembly installed on the
reservoir. Starting the HPS with the solid cap (that is used for shipping) on the reservoir
instead of the air breather assembly could cause pump cavitation.
The reservoir holds the oil that is needed to operate the various hydraulic components. For approximate
reservoir capacity, and information on adding oil to or changing the oil in the reservoir, refer to
maintenance information provided in the System Operating Instructions (supplied separately). The
reservoir is equipped with a sight gauge, air breather assembly and screen trapper.
Shut down the HPS and discharge hydraulic pressure before disconnecting any hydraulic
fluid coupling.
Do not disconnect any hydraulic coupling without first shutting down the HPS and
discharging stored pressure to zero. Tie down or otherwise secure all pressurized hoses to
prevent movement during system operation and to prevent the hose from whipping about
in the event of a rupture.
Содержание 1000KPX-J3
Страница 16: ...Chapter 1 Introduction 16 M47 17036 EN This page is intentionally blank...
Страница 52: ...Chapter 3 Additional System Details 52 M47 17036 EN This page is intentionally blank...
Страница 56: ...Chapter 4 Parts Replacement 56 M47 17036 EN This page is intentionally blank...
Страница 57: ......
Страница 58: ...Product Support www instron com...