background image

 

SRC- C1    

 

Page 4 

Subject to alteration 

Configuration parameters

 

The transmitter can be adapted to fit perfectly into any application by adjusting the software parameters. The 

parameters are set with the operation terminals OPA-S. The OPA-S may also be used as remote indicator.  

For correct display version 1.4 of OPA-S is required.  

Input configuration

 

Parameter

 

Description

 

Range  

Default  

IP 00 

Enable display of air quality LED on front.  

ON, OFF 

ON 

IP 01 

Samples taken for averaging control signal 

1…255 

10 

IP 02 

Calibration 

-10…10% 

IP 03 

Minimum CO2 range ppm  

(concentration when output is at its minimum.  

0…9900 ppm 

0 ppm 

IP 04 

Maximum CO2 range ppm  

concentration when output is at maximum) 

0…9900 ppm 

2000 ppm 

IP 05 

Level for medium air quality (orange light) 

0…9900 ppm 

800 ppm 

IP 06 

Level for high air quality (red light) 

0…9900 ppm 

1500 ppm 

Output configuration

 

Parameter

 

Description

 

Range  

Default  

OP 00 

AO1: CO2: Configuration of output signal: 

0 = Feedback CO2 input, 

1 = Feedback CO2 minimum value 

2 = Feedback CO2 maximum value 

0 – 2 

OP 01 

AO1: CO2: Minimum limitation of output signal 

0 – Max % 

0% 

OP 02 

AO1: CO2: Maximum limitation of output signal 

Min – 100% 

100% 

Calibration

 

The default sensor OEM unit is maintenance free in normal environments thanks to the built-in self-correcting ABC 

algorithm (Automatic Baseline Correction). This algorithm constantly keeps track of the sensor’s lowest reading over a 

7 days interval and slowly corrects for any long-term drift detected as compared to the expected fresh air value of 400 

ppm CO2. 

Rough handling and transportation might, however, result in a reduction of sensor reading accuracy. With time, the 

ABC function will tune the readings back to the correct numbers. The default “tuning speed” is however limited to 

about 30 ppm/week. For post calibration convenience, in the event that one cannot wait for the ABC algorithm to cure 

any calibration offset, jumper 3 is provided for the operator to choose calibration options.  

 

If jumper 3 is placed on position 1-2, for a minimum time of 8 seconds, the internal zero calibration is executed, in 

which case the sensor must be purged by some gas mixture free from CO2 (i.e. Nitrogen or Soda Lime CO2 scrubbed 

air). The air quality LED is blinking in green color alternately with the status LED in 2 second intervals.  

If jumper 3 Is placed on position 3-4, it is assumed that the sensor is operating in a fresh air environment (400 ppm 

CO2). The air quality LED is blinking in red color alternately with the status LED in 2 second intervals.  

 

If unsuccessful, please wait at least 1 minute before repeating the procedure again. Make sure that the sensor 

environment is steady and calm!  

 

Make sure to return jumper 3 to its default position of 2

- 3!   

 

Zero calibration procedure

 

The zero calibration procedure is as follow: 

1.

 

Connect the sensor on top with a tube (soft tubing 2x4 mm) and a 

nipple (nylon tubing 30x0.8x2.2 mm), see Figure 4 below. There 

are 2 alternative positions for nipple attachment. 

2.

 

Let a gas mixture flow into the sensor through the applied tube. 

The flow shall be in the range of 0.3 – 1.0 liter/minute during 3 

minutes. Keep the gas mixture flowing during the whole procedure. 

3.

 

Set jumper 3 to position 1-2 for a minimum time of 8 seconds. The 

air quality LED will blink in green color alternating with the status 

led in 2 second intervals.  

4.

 

Verify the zero calibration using the OPA-S or the analog outputs. 

They should show 0 ppm CO2. 

5.

 

If zero calibration is not executed (sensor detected unstable gas 

concentration) wait 10 sec and repeat steps 3 and 4 again. Do not 

breathe on the sensor! 

 

Insight  Control

Ph 1300 665 831

www.insightcontrol.net.au

Содержание SRC-C1

Страница 1: ... Indication of air quality A three color LED is used to indicate air quality Green light for low CO2 concentration orange light for medium and red light for high concentration The levels for low medium high may be programmed Default settings are 0 ppm low 800 ppm medium 1500 ppm high Use as P controller Set a minimum concentration when the extract fan should start to run at its minimum speed For e...

Страница 2: ...ion Climatic Conditions Temperature Humidity To IEC 721 3 3 class 3 K5 0 50 C 32 122 F 95 R H non condensing Transport Storage Climatic Conditions Temperature Humidity Mechanical Conditions To IEC 721 3 2 and IEC 721 3 1 class 3 K3 and class 1 K3 30 70 C 22 158 F 95 R H non condensing class 2M2 Standards conformity EMC Directive Low Voltage Directive 2004 108 EC 2006 95 EC Product standards Automa...

Страница 3: ...4 20mA 3 2 1 3 2 1 Status LED LED indicates status No light no power or unit damaged 5 sec blinking Normal 2 sec blinking OPA S connected 1 sec blinking Sensor element error if alternate in 2 sec interval with air quality LED Calibration is active OPA S 1 2 3 JP1 JP2 STATUS JP3 1 2 3 4 Calib 0 400 ppm G 24V G0 0V U1 I1 AO CO2 JP3 Calibration Calibrate 0 ppm Provide CO2 free gas Calibrate 400 ppm P...

Страница 4: ...tuning speed is however limited to about 30 ppm week For post calibration convenience in the event that one cannot wait for the ABC algorithm to cure any calibration offset jumper 3 is provided for the operator to choose calibration options If jumper 3 is placed on position 1 2 for a minimum time of 8 seconds the internal zero calibration is executed in which case the sensor must be purged by some...

Отзывы: