MEASURE OF SUCCESS

Intelligent, Economy

2-wire Transmitter

86BMQ00007 2009-11

GENERAL INFORMATION

MinIPAQ-HLP is a basic, programmable, non-isolated, easy-to-use 

2-wire  temperature  transmitter  intended  to  be  used  in  industrial 

environment. 

The transmitter is configured from a PC via a USB port, connected 

via the INOR USB Interface, by using the ConSoft program together 

with the MinIPAQ Soft program.  

When the transmitter is set from a PC no calibration is needed. 

ConSoft is compatible with Windows 2000 (SP3), Windows XP and 

Windows Vista (32-bit).

All necessary hardware and software for configuration are included 

in the ICON Configuration Kit, 70CFGUS001.

USER INSTRUCTIONS

The user instruction must be read prior to adjustment and/or installation. 

All information subject to change without notice.

MEASURE OF SUCCESS

This product should not be mixed with other kind of scrap, after usage.

It should be handled as an electronic/electric device.

www.inor.com, www.inor.se

INOR Process AB

, PO Box 9125, SE-200 39 Malmö, Sweden, 

Phone: +46(040)31 25 60, Fax: +46(040)31 25 70, E-mail: [email protected]

INOR Transmitter OY

, Unikkotie 13, FI-01300 Vantaa, Finland, 

Phone:+358-10-4217900, Fax: +358-10-4217901, E-mail: [email protected]

INOR Transmitter GmbH

, Rodenbacher Chaussee 6, 

DE-63457 Hanau-Wolfgang, Germany, 

Phone: +49(0)6181-582940, Fax: +49(0)6181-582944, E-mail: [email protected]

INOR Transmitter Inc

, 7 Dearborn Road, Peabody, MA 01960, USA, 

Phone: +1(978) 826-6900/(262) 670-9050, 

Fax: +1(978) 535-3882/(262) 670-9001, E-mail: [email protected]

CONFIGURATION

Guidelines for ConSoft version 1.1.0.2 or higher. 

 

 

 

 

 

 

 

 

 

 

Before making a configuration of MinIPAQ-HLP you need to do  

following:

1.  Install the drivers for the INOR USB Interface.

2.  Install the PC configuration software ConSoft in your PC.

3.  Install the PC configuration software MinIPAQ Soft, version 1.24.00 or  

 

higher, in your PC.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Please refer to “Installation Guide for ICON”, 86BCONF004, for  

 

detailed installation instructions. All software and guides are included in the 

ICON Configuration Kit, 70CFGUS001, and can also be downloaded from the 

web site www.inor.com. 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Configuration procedure:

The help files in ConSoft and MinIPAQ Soft contains complete information 

about the configuration procedure. Here is a short-form guide:

1.  Connect the MinIPAQ-HLP to the PC via the USB Interface.  

 

 

Configuration can be performed with or without connected power    

 

supply.

2.  In ConSoft, click on the upper left “Read from transmitter” button.   

 

The software will automatically recognize the connected transmitter  

 

and open the MinIPAQ Soft program. The MinIPAQ Soft program    

 

will automatically find the correct communication port (first time    

 

MinIPAQ Soft is used it will open the Com port selection window). 

3.  In the configuration window you can set the following parameters:

 

a.  Input (RTD or Thermocouple type)

 

b.  3- or 4-wire connection (for RTD input)

 

c.  Activation of Cold Junction Compensation (for T/C input)

 

d.  Measuring range in three different engineering units

 

e.  Upscale or downscale action at sensor break or short-circuit

 

f.  Compensation for known sensor errors

 

g.  Activation of a Low pass filter

4.  The selected configuration is downloaded to the transmitter by clicking  

 

the button “Transfer to transmitter”. If the transmitter is in operation  

 

during transmission from the PC, the transmitter output is frozen.    

 

When the transfer is complete, the transmitter begins using the new  

 

parameters directly.

DATA (shortform)

Input 

See  ”Input Connections”

Sensor failure

 

Upscale, downscale or off

Adjustments - Zero

 

Any value within range limits

Adjustments - Minimum spans

Pt100, Pt1000, Ni100, Ni120, Ni1000  10°C / 18°F

Cu10 

100°C / 180°F

T/C 

2 mV

Output

 

4-20 mA, temperature linear

Operating temperature

 

-40 to +85°C / -40 to +185°F

Galvanic isolation

 

No

Power supply

 

8 to 32 VDC

Typical accuracy

 

±0.15 % of temperature span

Connection head

 

DIN B or larger