GENERAL INFORMATION
IPAQ R202 is a low cost, non-isolated, easy-to-use 2-wire temperature
transmitter for Pt100 input intended to be used in industrial environment.
IPAQ R202 is designed to be mounted on a 35 mm DIN Rail.
The transmitter is configured from a PC via a USB port, connected via the
INOR USB Interface, by using the ConSoft program.
All necessary hardware and software for configuration are included in the
ICON Configuration Kit, 70CFGUS001.
R202
DATA (short form)
Input Pt100
3-wire connection
Pt100 (IEC 60751,
α
=0.00385)
-200 to +850 °C / -328 to +1562 °F
Adjustments - Zero Any value within range limits
Adjustments - Minimum span 20 °C / 36 °F
Output
4-20 mA, temperature linear
Sensor failure indication
Upscale (≥21.0 mA) or downscale (≤3.6 mA)
NAMUR Compliance
Current limitations and failure currents acc.
to NAMUR NE 43
Response time
0.13 to 54 s, adjustable filtering level
Ambient temperature
Storage and operation
-40 to +85 °C / -40 to +185 °F
Galvanic isolation No
Power supply
6 to 35 VDC
Accuracy
Max. of ±0.1 °C / ±0.2 °F or ±0.1 % of span
Sensor error correction
Via two-point correction
Runtime counter
Hour counter for elapsed operational time
Mounting
DIN rail acc. to EN 60715 / EN 50022, 35 mm
Factory settings
Input
Pt100 3W, 0 to 100 °C
Output
4 to 20 mA
Filtering level
Response time 0.9 s
Sensor failure indication
Upscale (≥21.0 mA)
CONFIGURATION
Before making a configuration of IPAQ R202 you need to do
following:
1. Install the configuration software ConSoft and drivers for the
INOR USB communication interface on a PC.
2. Connect the USB Interface to a free USB port of your PC and
wait for automatic installation of the USB Interface.
Please refer to the installation guide for ConSoft for detailed
installation instructions.
Required versions of configuration software:
ConSoft, 3.1.0 or higher
USB Interface, firmware 1.2.06 or higher
Connections:
PC to Interface: through USB type A to USB type B connection cable
Interface to transmitter: Mini USB type B to Mini USB type B cable
Configuration procedure:
1. Connect the transmitter to the PC via the USB Interface, LED indicator
“DEV” on the USB Interface becomes green.
Configuration can be performed with or without connected power
supply to the transmitter.
2. In ConSoft, click on the “Read” button. The software will recognize
the connected transmitter and open the configuration window.
3. In the “C202/R202” configuration window you can edit:
a. Measuring range in three different engineering units
b. Upscale or downscale action at sensor errors
c. Compensation for known sensor errors
d. TAG-number
4. The selected configuration is downloaded to the transmitter by
clicking the button “Write”, when the data transfer is complete,
the transmitter begins using the new parameters directly.