96-06434-00-01 revA
Service Manual, Inogen One G3HF
-120-
SP-522 Inogen One G3 Column Receptacle Replacement
Parts Required:
1. SP-522, Kit, Column Receptacles
Tools Required:
1. T10 Torx Driver
Notes:
1. When removing screws from the device carefully set in a bin or secure area as these
screws will be used to secure the replacement parts.
2. Many of the screws used on the Inogen One G3 have pre-applied loctite on the threads.
These loctite screws can be re-used up to four times before replacement screws are
required (If exceeded four uses please contact Inogen for a screw replacement kit).
Instructions for Removing the Housing:
1. While facing the front housing, remove the battery from the concentrator by pushing
down on the blue battery release clip and sliding the device away (Figure 1).
Figure 1: Direction to slide device to remove battery
2. To remove the front housing, place the device on its back side and remove the two screws
that secure the front housing to the chassis (Figure 2).