96-06434-00-01 revA
Service Manual, Inogen One G3HF
-102-
SP-558 Inogen One G3HF Tube Bracket and Muffler Replacement
Parts Required:
1. SP-558, G3HF Tube Bracket and Muffler
Tools Required:
1. T10 Torx Driver
Notes:
1. When removing screws from the device carefully set in a bin or secure area as
these screws will be used to secure the replacement parts.
2. Many of the screws used on the Inogen One G3 have pre-applied loctite on the threads.
These loctite screws can be re-used up to four times before replacement screws are
required (If exceeded four uses please contact Inogen for a screw replacement kit).
Instructions for Removing the Housing:
1. While facing the front housing, remove the battery from the concentrator by
pushing down on the blue battery release clip and sliding the device away (Figure
1).
Figure 1: Direction to slide device to remove battery