INV-430J/PE   
 
 
 
 
 
 

 

 

Finger Pulse Oximeter 

 

User Manual 

 
 
 

Innovo Groups 

406 Sterling Street, Building 2 

Camphill, PA, 17011 

 
 

Содержание INV-430J/PE

Страница 1: ...ps 406 Sterling Street Building 2 Camphill PA 17011 Phone 1 858 888 9781 cs innovogroups com http www innovogroups com INV 430J PE Finger Pulse Oximeter User Manual Innovo Groups 406 Sterling Street Building 2 Camphill PA 17011 ...

Страница 2: ...by any means without prior written consent of Innovo Groups All statements information and recommendations in this document are provided AS IS without warranties guarantees or representations of any kind either express or implied The information in this document is subject to change without notice Innovo Groups reserves the right of final interpretation of this document Version 1 2 Released on May...

Страница 3: ...grade IP22 Contents Letter to Users 1 Standard Packing List 2 Chapter 1 Precautions Warnings and Symbol 2 1 1 Precautions 2 1 2 Warnings 5 1 3 Symbol 7 Chapter 2 About This Product 9 2 1 Overview 9 2 2 Working Principles and Usage 9 2 3 Product Features 10 2 4 Limitations 11 2 5 Innovative features of the INV 430J PE 12 Chapter 3 Product Structure Parameter Settings and Operation Steps 13 3 1 Sche...

Страница 4: ...tions 31 6 2 Safety Type 34 6 3 Storage and Transportation 35 33 6 Technical parameters Parameter Value Display range Oxygen saturation 35 to 100 Pulse rate 25 bpm to 250 bpm Resolution Oxygen saturation 1 Pulse rate 1 bpm Measurem ent precision Oxygen saturation 2 70 to 100 No requirement 69 Pulse rate 2 bpm Alert range Oxygen saturation Upper limit 50 to 100 Lower limit 50 to 100 Pulse rate Uppe...

Страница 5: ...ple technological innovations you will come to see that the INV 430J PE is one of the most user friendly reliable and comfortable pulse oximeters that you will ever come to own Before using this product please carefully read the manual and follow the instructions stated herein Check whether all accessories are complete against the packing list and whether any component is damaged during shipping I...

Страница 6: ...ion of your finger is affected during prolong use This product is not designed to be used by newborn babies Seek for medical care if the measured value goes 31 Chapter 6 Technical Description and Safety Type 6 1 Technical Specifications 1 Dimensions 62 0 mm Width 37 0 mm Depth 32 0 mm Height Weight 42 5 g including the height of the two AAA dry batteries 2 Peak wavelength range of the light emitte...

Страница 7: ...please do not stare at the light emitting components of the Oximeter as that could cause harm and or potentially blind your eyes This pulse oximeter is not a medical device and is not intended to diagnose or treat any medical condition or disease It is intended for non medical use in healthy people to monitor their pulse and blood oxygen levels during sports and or aviation only People who need Sp...

Страница 8: ...e vascular atrophy severe anemia or low oxygen The user is in sudden cardiac arrest or shock state The user is wearing nail polish or artificial nails 29 Symptom Possible Cause Solution The displayed value of the oxygen saturation or pulse rate fluctuates 1 The finger is not placed properly between the sensor and the OLED 2 The user is moving his finger or body 1 Insert the finger again 2 Ensure t...

Страница 9: ...ot use the Oximeter in an environment with any flammable gases flammable anesthetic or other flammable substances Warning Keep unit and lanyard away from children as the included lanyard may present an entanglement or choking hazard to small children Adult supervision required never leave children unattended with unit or lanyard Warning Do not throw the batteries into fire as that could cause an e...

Страница 10: ...intenance Remove the batteries from the battery slot and properly store them if you do not plan to use the Oximeter for a long period of time Replace the batteries if they are low on power Clean the Oximeter and the fingertip before use to ensure accurate reading Store the Oximeter between 14 and 122 F 10 to 50 C and at humidity levels no greater than 93 Periodically check the Oximeter for damage ...

Страница 11: ... soft cloth dipped with 75 medical alcohol Clean the finger to be measured using the medical alcohol for disinfection purposes before and after use Do not clean the instrument by using high temperature high pressure disinfecting gas 7 1 3 Symbol Symbol Meaning BF type application part Caution Please see this manual SpO2 Symbol of Oxygen saturation bpmPR Symbol of Pulse rate Manufacture information...

Страница 12: ...4 Cleaning and Disinfection 4 1 Cleaning Ensure that the appearance of the instrument is clean To clean the instrument power off and remove the batteries first Then clean the outer surface of the instrument including the OLED screen using a piece of dry soft cloth dipped with 75 medical alcohol Do not immerse the unit in alcohol Caution Do not use any strong dissolving agent such as acetone Cautio...

Страница 13: ...meter involved in respiration and circulation The Pulse Oximeter feature herein is small portable non invasive and easy to use The user only needs to insert a finger into the chamber to measure his her SpO2 and pulse rate 2 2 Working Principles and Usage Oxygenated blood absorbs light at 660nm red light whereas deoxygenated blood absorbs light preferentially at 905nm infra red A pulse oximeter wor...

Страница 14: ... Low Battery indicator The device will automatically shut down if no signal is detected after 10 seconds The device is equipped with two 1 5 V AAA batteries which will allow the pulse oximeter to operate for approximately 30 hours 23 to keep your finger still during measurement It is also not advisable to use this instrument during sports activities as movement may lead to inaccuracies Once the re...

Страница 15: ...ence you will have to wait at least 3 6 seconds before a result will be displayed on the monitor If the pulse oximeter fails to detect blood flow for the first few seconds then the results will be delayed accordingly The pulse oximeter does not work for people with naturally small fingers or low peripheral blood flow This is because the pulse oximeter would not have enough blood to take a measurem...

Страница 16: ... cause discomfort 5 Using solid ABS plastic to block ambient light from reaching the sensor so that the signal to background ratio and the accuracy of the measurement is dramatically increased 6 Having a plethysmograph that allows the user to see and gauge the amount of blood detected by the pulse oximeter This will allow the user to know if the finger is properly inserted into the finger chamber ...

Страница 17: ...ng through the sling hole and thread the coarse end of the sling through the thin end of the sling and tighten the sling 13 Chapter 3 Product Structure Parameter Settings and Operation Steps 3 1 Schematic Structural Diagram OLED screen Power on key Battery cover ...

Страница 18: ...he power on function key to enter into the menu interface PRESSING 0 5 sec on the power on function key will allow you to to scroll through options while HOLDING 0 5 sec it will select an item indicates the option you are currently at Please see 19 3 4 Operation Steps 3 4 1 Installing the Batteries Install the two AAA dry batteries into the battery slot according to polarity indication and mount t...

Страница 19: ...nction key to decrease the upper or lower limit Move to the Exit option and hold the function key to return to the monitoring interface 15 figure below 3 3 2 Alarm Sound Setting To turn the alarm on off PRESS the power on key function key to move to Alm HOLD the power on key functional key to switch the option from on to off or vice versa When Alm is set to on and the measured values of the blood ...

Страница 20: ... in the plethysmograph see above corresponds to a heartbeat and the wave amplitude corresponds to the amount of blood detected by the pulse oximeter flowing through the blood vessels The pulse oximeter is optimized when the height of the wave amplitude is consistent throughout as shown in the figure above That is when you should take the reading The pulse oximeter can function with a PI reading as...