1-8
For Safe Operation
1
Flat Panel Sensors Operation Manual 897N200608
1.2 Electromagnetic Compatibility (EMC)
1.2.1 RIC
This equipment has been tested and found to comply with the limits for medical devices to the IEC
60601-1-2 (EN 60601-1-2), Medical Device Directive 93/42/EEC.
These limits are designed to provide reasonable protection against harmful interference in a typical
medical installation.
This equipment generates, uses and can radiate radio frequency energy and, if not installed and
used in accordance with the instructions, may cause harmful interference to other devices in the
vicinity.
However, there is no guarantee that interference will not occur in a particular installation.
If this equipment does cause harmful interference to other devices, which can be determined by
tuning the equipment off and on, the user is encouraged to try to correct the interference by one or
more of the following measures:
• Reorient or relocate the receiving device.
• Increase the separation between the equipment.
• Connect the equipment into an outlet on a circuit different from that to which the other device(s)
are connected.
If the problem cannot be solved with the above measures, stop using this equipment and consult
the manufacturer, our official dealer or INNOLUX Representative for help.
WARNING
● Do not place devices generating electromagnetic wave near this equipment.
● If a device(s) other than those specified is connected, predetermined EMC performance
cannot be guaranteed.