19
U S I N G T H E A N A LY Z E R
W A R M - U P
When power is first turned on, the analyzer will beep 5 times to indicate normal start-up, and then
enter the START-UP phase.
During WARM-UP, purge any residual gases from the sample hose by providing Zero air or Nitrogen to
the Sample Inlet. No other modes of operation can be initiated during WARM-UP. When complete, the
Measurement Screen is displayed. To configure the analyzer the “Menu” button is pressed from the
Measurement Screen and the Configuration Menu will be displayed.
In the MEASURE mode, the analyzer is processing the detector signals, converting them to
concentrations and displaying these measurements as well as sending data to the other outputs. The
following screen is displayed:
Z E R O I N G
The operator may initiate the ZERO cycle from the Home Screen. The analyzer will prompt for the type
of zero gas being used. Zero gas show be flowing throughout this process.
The options are Nitrogen, Argon or Air. Once the zero gas is selected the analyzer will start the zero
process and displays:
The process takes under 2 minutes, and the display will return to the Home Screen.