12
C A L I B R A T I O N
By pressing the
Calibration
button from
the main settings screen the analyzer goes
into standby and prompts for the type of
purge gas. The options are Nitrogen,
Argon and Air. Zero gas should flow
through the unit while zero is performed.
The analyzer goes into standby and begins
to purge the gas lines. The bar at the
bottom of the screen shows the progress
of the process as it moves from left to
right.
Once the “zeroing” has completed you will
see this screen. Enter each gas tag value
by pressing the corresponding box.
When entering the tag values you should
enter them just as they are on the
calibration bottle*.
Enter zero for the tag value of any gas that
you are not calibrating or for any gas not
present in the calibration bottle. Fields
containing a decimal point require tag
values to be entered as percentages.
Fields without a decimal point should be
entered as ppm.
*Use + and - to change the number that is
currently highlighted. To move to the
next digit or field use the <- or -> buttons.
Press the NEXT button to move on to the
next gas.
Please Select Zero Gas
N2/Ar
Air