Use
r’s
Manual of OMNI Express Patient Monitor
- 67
037.007.0001 Rev.2
SAFETY
CLASSIFICATION
◆
According to the degree of safety of application in the presence of FLAMMABLE
ANAESTHETIC MIXTURE WITH AIR, OR WITH OXYGEN OR NITROUS OXIDE:
The ISA is not suitable for use in the presence of FLAMMABLE ANAESTHETIC
MIXTURE WITH AIR, OR WITH OXYGEN OR NITROUS OXIDE.
◆
According to the degree of protection against harmful ingress of water: IPX4
◆
According to sterility: The ISA system contains no sterile parts.
◆
According to the model of operation: CONTINUOUS OPERATION
◆
According to the degree of protection against electric shock:
Nomoline Family sample lines are classified as DEFIBRILLATION PROOF TYPE BF
APPLIED PART
◆
The combination of OMNI Express and ISA shall be considered a ME SYSTEM.
WARNINGS
◆
The ISA sidestream gas analyzer is intended for use by authorized and trained
medical personnel only.
◆
Use only Nomoline sampling lines manufactured by PHASEIN.
◆
The ISA sidestream gas analyzer must not be used with flammable anesthetic
agents.
◆
Carefully route the sampling line to reduce the risk of patient entanglement or
strangulation.
◆
DO NOT re-use disposable sampling lines.
◆
DO NOT lift the ISA/OMNI Express by the sampling line as it could disconnect from
the ISA/ OMNI Express >, causing the ISA/ OMNI Express to fall on the patient.
◆
Used disposable sampling lines shall be disposed of in accordance with local
regulations for medical waste.
◆
DO NOT use adult/pediatric type sampling line configurations with infants, as this
may add dead space to the patient circuit.
◆
DO NOT use infant type sampling line configurations with adults, as this may cause
excessive flow resistance.
◆
DO NOT use the ISA sidestream gas analyzer with metered-dose inhalers or
nebulized medications as this may clog the bacteria filter.
◆
Check that the gas sample flow is not too high for the present patient category.
◆
Since a successful zeroing requires the presence of ambient air (21% O
2
and 0%
CO
2
) in the gas analyzer, ensure that the ISA is placed in a well ventilated place.
Avoid breathing near the ISA sidestream gas analyzer before or during the zeroing
procedure.
◆
The Nomoline sampling line and its interfaces are non-sterile devices. To avoid
damage, do not autoclave any part of the sampling line.
◆
Never sterilize or immerse the ISA sidestream gas analyzer in liquid.
◆
Measurements can be affected by mobile and RF communications equipment. Make
sure that the ISA sidestream gas analyzer is used in the electromagnetic environment
specified in this manual.
◆
ISA sidestream gas analyzer is intended only as an adjunct in patient assessment. It
must be used in conjunction with other assessments of clinical signs and symptoms.
◆
Replace the sampling line if the sampling line input connector starts flashing red, or a
Nomoline occlusion message is displayed on the OMNI Express.
◆
No modification of this equipment is allowed without authorization of the manufacturer.
If this equipment is modified, appropriate inspection and testing must be conducted to
ensure continued safe operation.
◆
ISA sidestream gas analyzers are not designed for MRI environments.
◆
During MRI scanning, the OMNI Express must be placed outside the MRI suite.
◆
Use of high frequency electrosurgical equipment in the vicinity of the ISA/OMNI
Express may produce interference and cause incorrect measurements.
◆
DO NOT use external ambient cooling of the ISA device.