Use
r’s
Manual of OMNI Express Patient Monitor
- 51
037.007.0001 Rev.2
alarm is activated; when the choice is
OFF
, the alarm indicator will not
light, the relative alarm parameter will not flash and relative parameter area will display
the
icon.
QTEMP UNIT
Fahrenheit or Celsius for choice, the factory setting is Celsius.
QTEMP ALARM UPPER-LIMIT
The QTemp alarm upper-limit, the range is
10
~
50
℃
(50~122
℉
)
, and the factory setting is
38.0
℃
(100.4
℉
)
, the single-step adjustable step-length is
0.1
℃
(0.2
℉
)
.
QTEMP ALARM LOWER-LIMIT
The QTemp alarm lower-limit, the range is
10
~
50
℃
(50~122
℉
)
, and the factory setting
is
36
℃
(96.8
℉
)
, the single-step adjustable step-length is
0.1
℃
(0.2
℉
)
.
MEASURE MODE
It provides nine measure modes as choice including Oral Fast, Oral Standard, Oral Cold,
Oral Direct, Oral Monitoring, Axillary Standard, Axillary Cold, Axillary Direct and Axillary
Monitoring.
MAINTAINENCE AND CLEANING
1. The entire device may be easily wiped clean. Water temperature should not exceed
130° F (55° C). Do not submerge or soak under water.
2. A mild detergent may be added to water. Use of cleaners such as Spray NineTM*,
PhisohexTM*, HibiclensTM*, or Vesta-SydeTM* CidexTM* may result in damage to
the thermometer case.
3. Use of 10:1 water and hypoclorite mixture or a damp isopropyl alcohol wipe
occasionally, is acceptable. Prolonged and repeated use of these chemicals may
result in damage to the thermometer case and display area.
4. Use of a cloth or sponge is recommended for cleaning. Abrasive pads may result in
damage to the thermometer case and display area.
5. This thermometer is provided non-sterile. DO NOT use ethylene oxide gas, heat,
autoclave, or any other harsh methods to sterilize this unit.
6. Isolation chambers may be replaced inexpensively instead of cleaning.
7. After cleaning the unit, shake the probe handle to drain out any excess solution.
Thoroughly dry the electrical contacts on both probe and thermometer.