Use
r’s
Manual of OMNI Express Patient Monitor
- 46
037.007.0001 Rev.2
SAFETY AND WARNING
1. Read this booklet thoroughly before using the FILAC 3000 electronic thermometer.
2. Do not use this device near flammable anesthetics. Not suitable for use in the
presence of flammable anesthetic mixture with air, oxygen or nitrous oxide.
3. Do not use this thermometer without first installing a new FILAC 3000 electronic
thermometer probe cover.
4. Use only FILAC 3000 electronic thermometer probe covers with this device. Use of
any other probe cover will result in erroneous temperature readings.
5. The device and probe covers are Non-sterile. Do not use on abraded tissue.
6. To limit cross contamination, use Blue devices for Oral and Axillary temperature
taking only.
7. Use RED devices only for RECTAL temperatures.
8. Thoroughly dry all electrical contacts on both probe and thermometer after washing,
or device may fail to function properly.
9. For re-calibration, service or integrity checks, refer to a qualified Biomedical
Technician or return to manufacturer.
10. Do not open unit. No user-serviceable parts inside. Opening of device may affect
calibration and voids warranty.
11. Disposal of used probe covers must be performed in accordance with current medical
practices or local regulations regarding disposal of infectious, biological medical
waste.
12. Removal of the batteries is recommended if the unit is not going to be used for an
extended period of time.
13. Dispose of batteries in a manner consistent with local environmental and institutional
policy for Lithium or Alkaline battery disposal.
14. Dispose of old battery-powered electronic equipment in a manner consistent with
institutional policy for expired equipment disposal.
15. Cleaning frequency and practices must be consistent with institutional policy for
cleaning of non-sterile devices.
16. Device to be used by trained personnel.
INSTRUCTION FOR USE
Probe Covers
—Applying & Removing
1. Open probe cover box by lifting tab at top corner and pulling
to remove top panel.
2. Insert box of probe covers into top of isolation chamber. (To
aid infection control, never switch boxes between blue and
red isolation chambers. Also, never switch probes between
blue and red isolation chambers. Keep like colors together.)
3. Remove probe from the probe well. This automatically turns
on the thermometer.
4. To help remind the user to apply or remove a probe cover, a
probe icon with flashing probe cover will be displayed when
the probe is withdrawn from the probe well and following a
completed temperature measurement.
[NOTE]:
Even though this device has been designed to minimize the effects of electromagnetic
interference, it does generate radio frequency energy. If not used in accordance with
the instructions, the device could cause interference in other equipment operating
within its vicinity. If the device is causing interference, the following actions may be
taken in an attempt to correct the interference
:
• Re-orient or re-locate the receiving device.
• Increase the separation between the devices.
•
Consult a customer service representative.