Use
r’s
Manual of OMNI Express Patient Monitor
- 30
037.007.0001 Rev.2
Resp Setup
Alarm Switch
Display
Resp Alarm High
Resp Alarm Low
Resp Gain
Sweep Speed
Sweep Color
Figure 14: Tree Diagram for Resp Menu
The menu has the following settings:
ALARM SWITCH
ON
and
OFF
for choice, the factory setting is
ON
.
If the RESP value is above or below the RESP alarm limit, when the choice is
ON
, the
alarm is activated; when the choice is
OFF
, the alarm indicator will not light, the relative
alarm parameter will not flash and relative parameter area will display the
icon.
RESP ALARM HIGH
The Resp alarm upper-limit, the range is
8
~
120
rpm. The default alarm limit is changed
with different patient type. When the patient type is adult or pediatric, the factory setting is
30
rpm; when it is neonatal, the factory setting is
100
rpm. The single-step adjustable
step- length is
1
rpm.
RESP ALARM LOW
The Resp alarm lower-limit, the range is
6
~
100
rpm
.
The default alarm limit is changed
with different patient type. When the patient type is adult or pediatric, the factory setting is
8
rpm; when it is neonatal, the factory setting is
30
rpm. The single-step adjustable step-
length is
1
rpm.
RESP GAIN
The user can freely choose one from items of X0.25, X 0.5, X1.0 and X2.0. The bigger the
gain is, the larger the waveform amplitude is. The factory setting is X1.0.
SWEEP SPEED
Choose from 6.25 mm/s, 12.5 mm/s and 25.0 mm/s, and the factory setting is 6.25 mm/s.
SWEEP COLOR
From White, Gray, Red, Yellow, Green, Cyan, Blue, and Magenta for choice, the default
setting is Cyan.
DISPLAY
The
ON
and
OFF
for choice
.
Pick
ON
can display Reap measured value, pick
OFF
would
not display the Resp value, but this do not influent the actual data of trend.
Applications
: when the patient’s thorax or abdomen is subjected too much interference,
the Resp monitoring is not accurate, so it is suggested to close the Resp display.
MAINTENANCE AND CLEANING
It is the same as the ECG monitoring.