3 - 32
XTC/3 Operating Manual
When XTC/3 is placed in the
Manual Power
mode by pressing the
MPWR
key,
power is affected by moving the
POWER/STOP
switch laterally. A
STOP
is
produced by pressing
POWER/STOP
.
A crystal switch is activated by pressing the red button on the body of the XTC/3.
This action alternates the active crystal of a dual head configuration or advances
the multi-position sensor to the next available crystal.
NOTE:
Upon leaving the
MANUAL POWER
state, XTC/3 enters the
DEPOSIT
state. The deposition will terminate if the
Final Thickness
parameter value
has been exceeded.
The Hand-held Controller includes a convenience hook for the controller that can
be attached to XTC/3 or other accessible location.
3.6.12 Lock Code
If a Lock Code has been entered in the
Process
sub-screen of the
General
Parameter
screen, the
L Lock
message will be displayed. If XTC/3 has been
locked via Remote Communications, the
R Lock
message will be displayed.
Parameters may be viewed but not changed from the front panel, unless the
lock code is reentered on the
Menu
screen while the cursor is positioned at
the
LOCK Code
selection.
A lost lock code may be cleared by pressing the
CLEAR
key immediately on
power-up and holding it briefly. User programmed parameters will be retained.
NOTE:
If no lock code is present, all parameters will be cleared.
3.6.13 Parameter Clear
If no Lock Code is present, all user-programmed parameters are cleared, except
for the IP Address, setting them to their default values, by pressing the
CLEAR
key
on power up. Repeat the process if a Lock Code was present.
3.6.14 Datalog
If the communication type is set to
Datalog
, the RS-232 port will become a "send
only" port (i.e., will not accept commands from a remote computer). When a layer
leaves deposit (either at the end of deposit or following a Stop), a stream of data
will be output. The data is a series of ASCII strings, each separated by a carriage
return and a line feed, in the order below:
1
Layer x
2
Film x (x = film number or the name if the name (XTC/3M only) is programmed)
3
Rate = xxx.x Å/s (0.0 - 999.9)
4
Thickness x Å (x = 0 - 999999 Angstroms)
5
Deposit Time = xx:xx Min:S