8 | Cleaning and maintenance
INFICON
104 / 140
UL3000-Fab-Operating-instructions-iina73en1-08-(1910)
8.4 Replacing the SL200 sniffer line filter
Structure of the sniffer
tip
Fig. 14: Sniffer tip
1 Capillary filter (plastic; standard)
4 Cross-head screws
2 Felt pads
5 Sinter filter
3 Capillary filter (metal with gasket;
option)
6 Gasket
A blockage of the sniffer tip can have the following causes:
Capillary filter blocked -> replace felt discs or capillary filter
Sinter filter clogged -> check or replace sinter filter
Capillary clogged in sniffer tip -> replace sniffer tip
Sniffer tip damaged -> replace sniffer tip
Sniffer line damaged / blocked -> replace SL 200.
Replace felt discs or
capillary filters
• Switch off the leak detector or disconnect the sniffer line from the device.
• Use tweezers to remove the dirty felt discs and replace them with new ones.
or
Содержание ULTRATEST UL3000 Fab
Страница 2: ...INFICON GmbH Bonner Strasse 498 50968 Cologne Germany ...
Страница 111: ...INFICON Decommissioning the measuring instrument 9 UL3000 Fab Operating instructions iina73en1 08 1910 111 140 ...
Страница 132: ...11 Appendix INFICON 132 140 UL3000 Fab Operating instructions iina73en1 08 1910 11 2 CE Declaration of Conformity ...
Страница 133: ...INFICON Appendix 11 UL3000 Fab Operating instructions iina73en1 08 1910 133 140 ...
Страница 134: ...11 Appendix INFICON 134 140 UL3000 Fab Operating instructions iina73en1 08 1910 ...
Страница 137: ...INFICON Index UL3000 Fab Operating instructions iina73en1 08 1910 137 140 Z ZERO 74 Button 23 Factory setting 67 ...
Страница 138: ...Index INFICON 138 140 UL3000 Fab Operating instructions iina73en1 08 1910 ...
Страница 139: ......
Страница 140: ......