10 – 4
Transpector SPS Operating Manual
10.4 RGA Configuration—Transpector SPS Tab
The Transpector SPS system must be configured before use. This section explains
the
Transpector SPS
tab in detail, but not the additional tabs in the
RGA
Configuration
window. Additional information is located in the
FabGuard Explorer
Operating Manual
(PN 074-528-P1).
10.4.1 Valves and Orifices Pane
The orifice in the inlet system is replaceable. Transpector SPS is set up in the
factory with the correct orifice value. However, if an orifice is changed, the new
configuration must be entered in the RGA Configuration.
10.4.1.0.1 Inlet Calibration
Defines the V5 Calibration valve.
If a standard calibration gas is installed, select
Ar standard
.
If a high mass tuning reference gas is installed, select
FC5311 standard
.
If V5 is not installed, select
No Valve
.
Figure 10-5 Inlet calibration
10.4.2 Gauges Pane
10.4.2.1 Process Gauge
Transpector SPS includes an optional CDG (capacitance diaphragm gauge).
If the CDG is installed, enter its appropriate pressure range. Note that the scale
is in millitorr.
If the CDG is not installed, clear the check box.
Figure 10-6 Process gauge