SQM-160 Operating Manual
5.4.16 Command: W
Parameters: . . . . . . . . . . . . . . . . . . . None
Description: . . . . . . . . . . . . . . . . . . . Returns the rate, thickness & frequency for
each sensor simultaneously.
Example: . . . . . . . . . . . . . . . . . . . . . W 0.0 00.78 58.654 5497324.373 00.78
58.668 5497433.978 00.78 58.648
5500330.091 00.79 58.654 5500440.187
00.77 58.655 5500550.287 00.78 58.662
5500660.020
In order: ignore the first 0.00 in the
response. Next: Sensor 1 Rate (Å/s), Sensor 1
Thickness (Å), Sensor 1 Freqency (Hz), Sensor 2 Rate
(Å/s) Sensor 2 Thickness (Å), Sensor 2 Frequency
(Hz), etc.
5.4.17 Command: Y
Parameters: . . . . . . . . . . . . . . . . . . . None.
Description: . . . . . . . . . . . . . . . . . . . Read the Power-Up Reset flag. The
Power-Up Reset flag is set during boot-up of
the SQM-160 and stays set until read
through the RS-232 interface. After the flag is
read, it is reset and will not be set again until
the SQM-160 is power cycled.
Example: . . . . . . . . . . . . . . . . . . . . . Y A1
Power-Up Reset flag is set.
Y A0
Power-Up Reset flag is reset.
5.4.18 Command: Z
Parameters: . . . . . . . . . . . . . . . . . . . None.
Description: . . . . . . . . . . . . . . . . . . . Set all Film and System menu parameters to
defaults. Note that this command can take
over 1 second to complete
Example: . . . . . . . . . . . . . . . . . . . . . Z A
All Film and System parameters are set to
defaults.
Содержание SQM-160
Страница 1: ...O P E R A T I N G M A N U A L SQM 1 60 Multi Film Rate Thickness Monitor IPN 074 511 P1C ...
Страница 2: ......
Страница 5: ......
Страница 6: ......
Страница 8: ......
Страница 22: ...1 10 IPN 074 511 P1C SQM 160 Operating Manual This page is intentionally blank ...
Страница 30: ...2 8 IPN 074 511 P1C SQM 160 Operating Manual This page is intentionally blank ...
Страница 60: ...4 12 IPN 074 511 P1C SQM 160 Operating Manual This page is intentionally blank ...
Страница 76: ...5 16 IPN 074 511 P1C SQM 160 Operating Manual This page is intentionally blank ...
Страница 98: ...6 22 IPN 074 511 P1C SQM 160 Operating Manual This page is intentionally blank ...
Страница 110: ...8 8 IPN 074 511 P1C SQM 160 Operating Manual This page is intentionally blank ...
Страница 120: ...A 10 IPN 074 511 P1C SQM 160 Operating Manual This page is intentionally blank ...