11 | Calibration
INFICON
38 / 80
Sensistor_Sentrac_Strix_Operating_Manual_ninb65en1-01-(2010)
• Relevant Certificate
For more information see Spare Parts and Accessories.
11.2 How to Calibrate
INFICON support two types of Calibration Leak. The one with large leaks must be
connected to tracer gas pressure to work. The small leak is connected to a small
fillable gas container. Read the instruction delivered with the leak how to set up and
how to use the leak.
11.2.1 Prepare the Calibration Leak
Leak with container:
1. Fill the gas container to the indicated pressure with tracer gas.
2. Leak is ready to use.
Leak without container:
1. Connect the leak with tracer gas with the pressure according the certificate.
2. Vent the hose from air via the Purge Valve on the leak.
3. Start to use the leak.
11.2.2 Set the Calibration Value
Calibrate
Setup 1
Info
Interval
Setup 2
Keep probe in background air
Settings > Calibration
1. Click Settings > Calibration > Setup 1.
2. Set the Calibration Unit and Calibration Value (as stated on the calibration
certificate issued for the leak).
3. If using Calibration Leak: Click Settings > Calibration > Setup 2 to set calibration
Leak Gas type.
4. Set Leak Gas (as stated on the calibration certificate issued for the leak)