Maintenance
97
7
Maintenance
7.1
Maintenance schedule
Required tools
• 2 screw drivers, size 2
• 1 ring spanner wrench, 19 mm
• 8 mm hexagonal screw driver (delivered with Protec P3000).
Key for Maintenance Schedule:
• I Repair level I Customer
• II Repair level II Customer with technical training from INFICON
• III Repair level III INFICON service engineer
1) Depends on the environment
For all maintenance on the Protec P3000, the main power must be
disconnected first.
Required
maintenance
Sub
Assembly
Material
Description
Part no.
Operation hours
Repair
level
2000 5000
10.00
0
Check sinter filter and
replace if necessary
Sniffer tip
Sinter Filter for
Sniffertip SL3xx,
SL3000-x (5 pcs.)
200 03 500 X
I
Replace if W35 „Flow
too low“
Sniffer tip
Felt for Capillary Filter
SL3xx,
SL3000-x (50pcs.)
200 001
116
1)
I
Check internal filter and
replace if necessary
Main Unit
internal Filter (10 pcs.)
200 001
680
X
II
Maintenance of
diaphragm pump
Diaphragm
Pump
MVP015
Wearing parts kit for
Diaphragm pump
200 03 504
X
III
Clean or replace the air
filter at the bottom of
the chassis
Main Unit
Air Filter Protec P3000
(104x154 mm; 5 pcs.)
200 001
552
X
I
Replace gas reservoir
after 1 year
PRO-Check
Replacement gas reservoir
for PRO-Check
521-010
1)
I
Replace filter pad
Sniffer tip
SL3000-XL
200 002
251
I
Содержание Protec P3000
Страница 2: ......
Страница 10: ...10 General Safety Precautions...
Страница 121: ...CE Declaration of Conformity 121 9 CE Declaration of Conformity Fig 91...
Страница 122: ...122 CE Declaration of Conformity...
Страница 125: ...Appendix 125...
Страница 126: ......
Страница 127: ......
Страница 128: ......