7 | Warning and error messages
INFICON
74 / 100
Pernicka-700H-CHLD-Operating-instructions-jin76aen1-04-(1911)
7 Warning and error messages
During operation, the display shows information that helps you operate the instrument.
Measurement values are displayed along with current device modes, operating
instructions as well as warnings and error messages. The instrument is equipped with
extensive self-diagnostic functions. If the electronics detect a faulty state, the device
will show this as far as possible on the display and will interrupt operation when
necessary.
Warnings
Warnings warn of instrument modes that can impair the accuracy of measurements.
Operation of the instrument is not interrupted.
Error messages
Errors are events that force the interruption of the operation. The error message
consists of a number and a descriptive text.
7.1 Display warning and error messages
► To open the "Errors" window, press the tab "Errors"."Errors" window"Errors"
windowOperationOperation
Fig. 48: Window "Errors"
In the "Errors" window, error messages are displayed during a measurement process
and recorded in the background.
Warnings
Message
Possible causes
Solution
Emission current > +/-
10 %
HV pressure is too
high.
Cathode is aged.
Perform a regeneration. See
72]".