6 | Installation
INFICON
128 / 319
074-594-P1H Micro GC Fusion Operating Man-
ual
19 Turn on Micro GC Fusion. The instrument is ready to perform an analysis.
CAUTION
If the flow rate through the device exceeds 100 mL/min, the output pressure
from the device to Micro GC Fusion may exceed the acceptable pressure
range, causing instrument damage.
WARNING
Safely vent carrier and sample streams - potentially toxic, noxious, asphyxiant
(oxygen displacing) or flammable gases - outside Micro GC Fusion and away
from the operating area. Vent toxic gases to a fume hood, chemical trap, or
reaction medium.
6.15.3 4-Module Micro GC Fusion Chassis Installation - Front
Install
The pressure reducer and gas-liquid separator includes the following items. These
items are needed to install the device on the front of a 4-module Micro GC Fusion
chassis.
PN
Description
Quantity
059-0696
5/16 in. female nut
1
059-0697
1/16 in. ferrule set
1
059-0336
1/16 in. Valco male nut
1
059-0335
1/16 in. ferrule
1
952-229-G1
sample tube
1
B3079191X
M3 screw
2
952-361-P5
bracket E
1
pressure reducer and gas-
liquid separator assembly
1
There are three tubes in kit (PN 952-229-G1) and the shortest length tube is
used for this installation.
Tools required:
• 3/8 in. open-ended wrench
• 5/16 in. open-ended wrench
• 1/4 in. open-ended wrench
Содержание Micro GC Fusion 2-Module
Страница 1: ...Operating Manual Micro GC Fusion Gas Analyzer 074 594 P1H ...
Страница 2: ...INFICON Two Technology Place East Syracuse NY 13057 9714 ...
Страница 13: ...INFICON Declaration of Conformity 2 074 594 P1H Micro GC Fusion Operating Manual 13 319 ...
Страница 318: ......
Страница 319: ...074 594 P1H 2019 ...