12.5.2 AutoTune Parameters
MATERIAL INDEX
. . . . . . . . . . . . . . 1 to 32
This parameter is used to select the Material that is to be AutoTuned. Values
range from 1 to 32. The default value is 1.
DESIRED RATE
. . . . . . . . . . . . . . . 0.001 to 999.9 Å/s
For a Complete Tune, this parameter determines the Rate for which the
AutoTuned control loop parameters will be calculated. For Quick Tune this
parameter is ignored. Values range from 0.001 to 999.9 Å/s. The default value
is 10.0 Å/s.
MAXIMUM RATE
. . . . . . . . . . . . . . . 0.002 to 999.9 Å/s
This parameter is used to set the maximum deposition Rate allowed during an
AutoTune session. The events which occur upon exceeding this value are
dependent upon whether a Quick Tune or a Complete Tune is designated. The
range of values permitted for this parameter is dependent on the value chosen
for the Desired Rate. The maximum allowed value will always be 999.9 Å/s.
The minimum allowed value is 2 times the Desired Rate value.
TUNE TYPE
. . . . . . . . . . . . . . . . . . . Quick/Complete
This parameter is used to choose between a Quick Tune and a Complete Tune
described in
.The default value is Quick.
MATERIAL NAME
This is a read-only parameter that identifies the material being deposited.
CONTROL LOOP
This is a read-only parameter that indicates the type of control loop specified
during material definition. It has no effect during Auto Tune. The type can be
Non-PID, PI or PID.
PROCESS GAIN
This is a read-only parameter that represents the rate of change of % source
power for a given rate deviation. It has no effect during Auto Tune.
TIME CONSTANT
This is a read-only parameter that represents the source's time constant. It is
valid only for a PI or PID control loop and has no effect during Auto Tune.
DEAD TIME
This is a read-only parameter that represents the source's lag time. It is valid
only for a PI or PID control loop and has no effect during Auto Tune.
Содержание IC6
Страница 1: ...O P E R A T I N G M A N U A L IC6 Thin Film Deposition Controller IPN 074 505 P1E Cover Page ...
Страница 2: ......
Страница 8: ......
Страница 20: ...TOC 12 IPN 074 505 P1E IC6 Operating Manual This page is intentionally blank ...
Страница 42: ...1 22 IPN 074 505 P1E IC6 Operating Manual This page is intentionally blank ...
Страница 134: ...6 6 IPN 074 505 P1E IC6 Operating Manual This page is intentionally blank ...
Страница 148: ...8 4 IPN 074 505 P1E IC6 Operating Manual This page is intentionally blank ...
Страница 170: ...9 22 IPN 074 505 P1E IC6 Operating Manual This page is intentionally blank ...
Страница 236: ...11 2 IPN 074 505 P1E IC6 Operating Manual Figure 11 1 The essential advantage of multi point deposition sensing ...
Страница 268: ...13 2 IPN 074 505 P1E IC6 Operating Manual This page is intentionally blank ...
Страница 312: ...15 40 IPN 074 505 P1E IC6 Operating Manual This page is intentionally blank ...
Страница 338: ...A 10 IPN 074 505 P1E IC6 Operating Manual This page is intentionally blank ...
Страница 342: ...1 4 IPN 074 505 P1E IC6 Operating Manual This page is intentionally blank ...