17 | Maintenance
INFICON
406 / 436
074-471-P1E HAPSITE ER Operating Manual
The high pressure error may also indicate that the NEG pump is worn and needs
replacing. Please continue reading for information on how to troubleshoot NEG
pumps.
To view the MS pressure:
1
Touch
HAPSITE
icon.
2
Touch the
Tune
icon.
3
Locate the MS pressure. In order for HAPSITE ER to properly operate, the
pressure must be below 6 e-03. If pressure is too high, ionizer will not activate.
If the NEG pump has 150 hours or less, consider trying a bakeout to extend the life of
the NEG pump. To check the NEG pump hours, refer to NEG Status [
191].
If the NEG pump has more than 250 hours, a bakeout can be tried, though the results
are likely to be limited.
17.5 Attaching HAPSITE to the Service Module
CAUTION
Prior to attaching HAPSITE ER to the Service Module, unplug the black
NEG cable inside the HAPSITE ER front panel. This will ensure that the
NEG will not heat when using the Service Module to provide vacuum.
WARNING
Venting HAPSITE ER with a NEG that has not cooled causes total NEG
consumption and may possibly result in severe damage to the HAPSITE
ER mass spectrometer components. It may also result in physical injury
since extreme heat generation from NEG consumption creates hot
surfaces.
If HAPSITE ER is turned on with the NEG at 400°C, then the NEG must be cooled
before proceding. Turn off HAPSITE ER and allow it to cool for approximately 24
hours or until the NEG is at room temperature.
Содержание HAPSITE ER
Страница 1: ...Operating Manual HAPSITE ER 074 471 P1E ...
Страница 2: ...INFICON Two Technology Place East Syracuse NY 13057 9714 ...
Страница 12: ...1 Declaration of Conformity INFICON 12 436 074 471 P1E HAPSITE ER Operating Manual 1 Declaration of Conformity ...
Страница 13: ...INFICON Declaration of Conformity 1 074 471 P1E HAPSITE ER Operating Manual 13 436 ...
Страница 56: ...9 Operate the Instrument in Portable Mode INFICON 56 436 074 471 P1E HAPSITE ER Operating Manual ...
Страница 184: ...12 ER IQ Software INFICON 184 436 074 471 P1E HAPSITE ER Operating Manual ...
Страница 208: ...13 Data Review INFICON 208 436 074 471 P1E HAPSITE ER Operating Manual 4 Double click on the desired data file ...
Страница 226: ...13 Data Review INFICON 226 436 074 471 P1E HAPSITE ER Operating Manual 5 Click Save 6 Click Run ...
Страница 237: ...INFICON Data Review 13 074 471 P1E HAPSITE ER Operating Manual 237 436 2 Check the Search in NIST box ...
Страница 238: ...13 Data Review INFICON 238 436 074 471 P1E HAPSITE ER Operating Manual 3 Click Search ...
Страница 248: ...13 Data Review INFICON 248 436 074 471 P1E HAPSITE ER Operating Manual 2 The report is displayed ...
Страница 250: ...13 Data Review INFICON 250 436 074 471 P1E HAPSITE ER Operating Manual 5 The report opens in Excel ...
Страница 275: ...INFICON Data Review 13 074 471 P1E HAPSITE ER Operating Manual 275 436 6 The identification is displayed on the screen ...
Страница 323: ...INFICON Method Editor 15 074 471 P1E HAPSITE ER Operating Manual 323 436 ...
Страница 330: ...15 Method Editor INFICON 330 436 074 471 P1E HAPSITE ER Operating Manual 2 Select Next until the SIM page is displayed ...
Страница 333: ...INFICON Method Editor 15 074 471 P1E HAPSITE ER Operating Manual 333 436 5 Enter the name of the chemical of interest ...
Страница 349: ...INFICON Method Editor 15 074 471 P1E HAPSITE ER Operating Manual 349 436 The Analyte List is displayed ...
Страница 371: ...INFICON Calibration 16 074 471 P1E HAPSITE ER Operating Manual 371 436 ð Click OK ...
Страница 381: ...INFICON Calibration 16 074 471 P1E HAPSITE ER Operating Manual 381 436 ...
Страница 389: ...INFICON Calibration 16 074 471 P1E HAPSITE ER Operating Manual 389 436 49 Click OK ...
Страница 390: ...16 Calibration INFICON 390 436 074 471 P1E HAPSITE ER Operating Manual 16 3 Definition of Terms in the Calibrate Window ...
Страница 435: ......
Страница 436: ......