11.2.2 Quick Reference SOP - Running Survey Mode
1
Double-click the
ER IQ
software icon.
2
Double-click the
Run Method
icon.
3
Wait for heaters to reach the setpoint temperatures.
4
Click the
RUN
button in the pop-up window.
5
Sample background for one minute and note the TIC.
6
Hold the probe over the sample until a response that is 2 to 3 times the
baseline is observed. If the TIC does not increase, sample for a full minute.
7
Press
STOP
to stop the method.
ð
This is a tentative identification. To confirm results, run an
Analyze
(GC/MS)
Method.
CAUTION
Do not touch the sample with the probe.
Do not allow liquids to enter the probe.
11.3 ANALYZE (GC/MS) Mode with the
Concentrator
11.3.1 Tri-Bed Concentrator
The Tri-Bed concentrator is used for analyzing samples with concentrations in the part
per million to high part per trillion range. Two default qualitative methods are ER_Tri-
Bed_PPM_Standard and ER_Tri-Bed_PPB Standard. Use the ER_Tri-
Bed_PPM_Standard method when a response is seen in Survey. If a compound is
suspected, but Survey does not show an increase in TIC (total ion count, which is a
measure of response), use the ER_Tri-Bed_PPB_Standard method.
CAUTION
The concentrator feature has increased sensitivity. Use the appropriate
method to avoid saturating HAPSITE ER.
Содержание HAPSITE ER
Страница 1: ...Operating Manual HAPSITE ER 074 471 P1E ...
Страница 2: ...INFICON Two Technology Place East Syracuse NY 13057 9714 ...
Страница 12: ...1 Declaration of Conformity INFICON 12 436 074 471 P1E HAPSITE ER Operating Manual 1 Declaration of Conformity ...
Страница 13: ...INFICON Declaration of Conformity 1 074 471 P1E HAPSITE ER Operating Manual 13 436 ...
Страница 56: ...9 Operate the Instrument in Portable Mode INFICON 56 436 074 471 P1E HAPSITE ER Operating Manual ...
Страница 184: ...12 ER IQ Software INFICON 184 436 074 471 P1E HAPSITE ER Operating Manual ...
Страница 208: ...13 Data Review INFICON 208 436 074 471 P1E HAPSITE ER Operating Manual 4 Double click on the desired data file ...
Страница 226: ...13 Data Review INFICON 226 436 074 471 P1E HAPSITE ER Operating Manual 5 Click Save 6 Click Run ...
Страница 237: ...INFICON Data Review 13 074 471 P1E HAPSITE ER Operating Manual 237 436 2 Check the Search in NIST box ...
Страница 238: ...13 Data Review INFICON 238 436 074 471 P1E HAPSITE ER Operating Manual 3 Click Search ...
Страница 248: ...13 Data Review INFICON 248 436 074 471 P1E HAPSITE ER Operating Manual 2 The report is displayed ...
Страница 250: ...13 Data Review INFICON 250 436 074 471 P1E HAPSITE ER Operating Manual 5 The report opens in Excel ...
Страница 275: ...INFICON Data Review 13 074 471 P1E HAPSITE ER Operating Manual 275 436 6 The identification is displayed on the screen ...
Страница 323: ...INFICON Method Editor 15 074 471 P1E HAPSITE ER Operating Manual 323 436 ...
Страница 330: ...15 Method Editor INFICON 330 436 074 471 P1E HAPSITE ER Operating Manual 2 Select Next until the SIM page is displayed ...
Страница 333: ...INFICON Method Editor 15 074 471 P1E HAPSITE ER Operating Manual 333 436 5 Enter the name of the chemical of interest ...
Страница 349: ...INFICON Method Editor 15 074 471 P1E HAPSITE ER Operating Manual 349 436 The Analyte List is displayed ...
Страница 371: ...INFICON Calibration 16 074 471 P1E HAPSITE ER Operating Manual 371 436 ð Click OK ...
Страница 381: ...INFICON Calibration 16 074 471 P1E HAPSITE ER Operating Manual 381 436 ...
Страница 389: ...INFICON Calibration 16 074 471 P1E HAPSITE ER Operating Manual 389 436 49 Click OK ...
Страница 390: ...16 Calibration INFICON 390 436 074 471 P1E HAPSITE ER Operating Manual 16 3 Definition of Terms in the Calibrate Window ...
Страница 435: ......
Страница 436: ......