10 | Communications and Touch Screen Options
INFICON
110 / 436
074-471-P1E HAPSITE ER Operating Manual
10 Communications and Touch Screen
Options
10.1 Communications
HAPSITE ER has two communication options: the cross-over cable and the wireless
connection. The wireless settings will be configured with the laptop at the factory.
Before connecting wirelessly, the wireless radio on HAPSITE ER and the wireless
button on the laptop will have to be enabled. Follow the instructions below to turn on
the radio. Refer to Connecting the Laptop [
42] for instructions on attaching the
laptop.
10.1.1 Wireless Range
HAPSITE ER is equipped with an 802.11 b/g wireless adapter. The typical range for a
single is 300 ft. (100 m) with no obstructions. The following may degrade the signal:
• Metal buildings
• Concrete structures
• Electric devices in the area
10.1.2 Turning On the Radio
This procedure gives instructions for turning on the radio, which is necessary for
wireless communication.
DANGER
When the HAPSITE ER radio is on, even if wireless communication is not
being used, a radio signal is being transmitted. The only way to eliminate
the radio signal is to turn the radio off.
1
Open the HAPSITE ER front panel.
2
Remove the cover from the power switch (for the wireless radio). It is located
on the far left side. To remove, unscrew the cover by turning it counter-
clockwise.
Содержание HAPSITE ER
Страница 1: ...Operating Manual HAPSITE ER 074 471 P1E ...
Страница 2: ...INFICON Two Technology Place East Syracuse NY 13057 9714 ...
Страница 12: ...1 Declaration of Conformity INFICON 12 436 074 471 P1E HAPSITE ER Operating Manual 1 Declaration of Conformity ...
Страница 13: ...INFICON Declaration of Conformity 1 074 471 P1E HAPSITE ER Operating Manual 13 436 ...
Страница 56: ...9 Operate the Instrument in Portable Mode INFICON 56 436 074 471 P1E HAPSITE ER Operating Manual ...
Страница 184: ...12 ER IQ Software INFICON 184 436 074 471 P1E HAPSITE ER Operating Manual ...
Страница 208: ...13 Data Review INFICON 208 436 074 471 P1E HAPSITE ER Operating Manual 4 Double click on the desired data file ...
Страница 226: ...13 Data Review INFICON 226 436 074 471 P1E HAPSITE ER Operating Manual 5 Click Save 6 Click Run ...
Страница 237: ...INFICON Data Review 13 074 471 P1E HAPSITE ER Operating Manual 237 436 2 Check the Search in NIST box ...
Страница 238: ...13 Data Review INFICON 238 436 074 471 P1E HAPSITE ER Operating Manual 3 Click Search ...
Страница 248: ...13 Data Review INFICON 248 436 074 471 P1E HAPSITE ER Operating Manual 2 The report is displayed ...
Страница 250: ...13 Data Review INFICON 250 436 074 471 P1E HAPSITE ER Operating Manual 5 The report opens in Excel ...
Страница 275: ...INFICON Data Review 13 074 471 P1E HAPSITE ER Operating Manual 275 436 6 The identification is displayed on the screen ...
Страница 323: ...INFICON Method Editor 15 074 471 P1E HAPSITE ER Operating Manual 323 436 ...
Страница 330: ...15 Method Editor INFICON 330 436 074 471 P1E HAPSITE ER Operating Manual 2 Select Next until the SIM page is displayed ...
Страница 333: ...INFICON Method Editor 15 074 471 P1E HAPSITE ER Operating Manual 333 436 5 Enter the name of the chemical of interest ...
Страница 349: ...INFICON Method Editor 15 074 471 P1E HAPSITE ER Operating Manual 349 436 The Analyte List is displayed ...
Страница 371: ...INFICON Calibration 16 074 471 P1E HAPSITE ER Operating Manual 371 436 ð Click OK ...
Страница 381: ...INFICON Calibration 16 074 471 P1E HAPSITE ER Operating Manual 381 436 ...
Страница 389: ...INFICON Calibration 16 074 471 P1E HAPSITE ER Operating Manual 389 436 49 Click OK ...
Страница 390: ...16 Calibration INFICON 390 436 074 471 P1E HAPSITE ER Operating Manual 16 3 Definition of Terms in the Calibrate Window ...
Страница 435: ......
Страница 436: ......