6 | Operation
INFICON
68 / 152
UL6000-Fab,-PLUS-Operating-instructions-iinc75en1-02-(2207)
6.3.8 Setting the machine factor
If you are measuring in the operation mode "Vacuum" and using an external pump in
parallel, the measured leak rate would be too small in comparison to the leak rate
based on an internal calibration.
If you would like to compensate for this you can multiply the measured leak rate with
the suitable machine factor and then have the value shown. This factor is used only in
the vacuum mode and not in the sniffing mode.
When taking into consideration the Helium pumping speed of the device, the machine
factor can also be estimated. The measurement of the leak rate is more accurate
using an external calibration leak with the test object - once with the external pump
switched and once without. The difference between the results determines the
machine factor.
The machine factor can also be used to correct the leak rate display with respect to
the air equivalence. The machine factor for this correction is 0.37.
ü
Operator
or
Supervisor
rights
ü
The operation mode "Vacuum" is set.
1
> Setup > Measurement > Calibration leak
ð
The field “Machine factor helium” is shown.
2
Adjust.
3
Save
.
6.3.9 Set external calibration leak
To use an external calibration leak for calibration, enter the leak rate of the calibration
leak.
ü
Operator
or
Supervisor
rights
1
> Setup > Measurement > Calibration leak
2
Adopt the printed value and the associated unit from the calibration leak or
certificate.
Do not change the combination of the printed value and associated unit, even if
your device is otherwise set to other units!
3
Save
.
See also
2
6.3.10 Change pressure limits
Pressure limits for the vacuum mode.
Содержание Fab UL6000 Fab
Страница 2: ...INFICON GmbH Bonner Strasse 498 50968 Cologne Germany ...
Страница 125: ...INFICON Decommissioning 9 UL6000 Fab PLUS Operating instructions iinc75en1 02 2207 125 152 ...
Страница 145: ...INFICON Appendix 11 UL6000 Fab PLUS Operating instructions iinc75en1 02 2207 145 152 11 2 CE Declaration of Conformity ...
Страница 146: ...11 Appendix INFICON 146 152 UL6000 Fab PLUS Operating instructions iinc75en1 02 2207 ...
Страница 147: ...INFICON Appendix 11 UL6000 Fab PLUS Operating instructions iinc75en1 02 2207 147 152 11 3 RoHS ...
Страница 150: ...Index INFICON 150 152 UL6000 Fab PLUS Operating instructions iinc75en1 02 2207 Factory setting 70 ...
Страница 151: ......
Страница 152: ......