4 | Description
INFICON
34 / 152
UL6000-Fab,-PLUS-Operating-instructions-iinc75en1-02-(2207)
UL6000 Fab, UL6000 Fab PLUS
Degree of contamination ***)
2
**) Measuring condition for 40°C ambient temperature: Measuring mode ULTRA, gas
load 35 sccm
***) According to DIN EN 61010-1: Only non-conductive contamination. Occasionally,
temporary conductivity may occur due to condensation.
4.5 Factory settings
Leak rate setpoint 1
1 x10
-9
mbar l/s
Leak rate setpoint 2
1 x10
-8
mbar l/s
Leak rate setpoint 3
1 x10
-7
mbar l/s
Leak rate setpoint 4
1 x10
-6
mbar l/s
Automatic purge
On
Volume
2
ZERO mode
I•ZERO 2.0
Calibration request
Off
TMP ventilation
On (can only be changed by service)
Interfaces leak rate unit vacuum operation
mbar l/s
Interfaces leak rate unit sniffer operation
mbar l/s
Interface pressure unit
mbar
Mass
4 (helium)
Protocol on the IO1000 RS232 connection
ASCII
Analog output IO1000 channel 1
Leak rate mantissa
Analog output IO1000 channel 2
Leak rate exponent
Analog output IO1000 scaling
0.5 V / decade
Digital output pin 1 (IO1000)
Calibration request (inverted)
Digital output pin 2 (IO1000)
Error or warning (inverted)
Digital output pin 3 (IO1000)
Setpoint 1 (inverted)
Digital output pin 4 (IO1000)
Setpoint 2 (inverted)
Digital output pin 5 (IO1000)
Measuring
Digital output pin 6 (IO1000)
ZERO active
Digital output pin 7 (IO1000)
Ready for operation
Digital output pin 8 (IO1000)
Calibration active
Digital input pin 1 (IO1000)
VENT
Digital input pin 2 (IO1000)
Sniff
Digital input pin 3 (IO1000)
START
Digital input pin 4 (IO1000)
Stop
Содержание Fab UL6000 Fab
Страница 2: ...INFICON GmbH Bonner Strasse 498 50968 Cologne Germany ...
Страница 125: ...INFICON Decommissioning 9 UL6000 Fab PLUS Operating instructions iinc75en1 02 2207 125 152 ...
Страница 145: ...INFICON Appendix 11 UL6000 Fab PLUS Operating instructions iinc75en1 02 2207 145 152 11 2 CE Declaration of Conformity ...
Страница 146: ...11 Appendix INFICON 146 152 UL6000 Fab PLUS Operating instructions iinc75en1 02 2207 ...
Страница 147: ...INFICON Appendix 11 UL6000 Fab PLUS Operating instructions iinc75en1 02 2207 147 152 11 3 RoHS ...
Страница 150: ...Index INFICON 150 152 UL6000 Fab PLUS Operating instructions iinc75en1 02 2207 Factory setting 70 ...
Страница 151: ......
Страница 152: ......