5 Installation
5.1 Transport protection
NOTICE
Property damage due to transport protection that has not been removed
The transport protection blocks the mechanical system in the Gas Detection Unit.
► Remove the transport protection before startup.
The transport protection is on the bottom of the Gas Detection Unit and consists of a
yellow star screw.
1
1
Transport protection
5.2 Setup
WARNING
Danger from moisture and electricity
Moisture entering the battery leak detector can lead to personal injury due to electric
shocks, as well as damage to property due to short circuiting.
► Only operate the battery leak detector in a dry environment.
► Operate the battery leak detector away from sources of liquid and moisture.