Maintenance
8-12
ki
n
a
22
e c
h
ap
te
r 8
.fm
te
chn
ica
l ha
nd
boo
k
(1
20
6)
3
Remove the sinter filter with the o-ring.
4
Visually check the filter for contamination.
5
Install a new sinter filter with o-ring at the bottom of the filter tip.
6
Reinstall the sniffer tip.
7
Switch on the Ecotec E3000.
8
Hold finger against the capillary filter:
You should be able to feel the resulting vacuum.
9
If the warning “Flow through capillary too low” remains after releasing the finger from
the capillary filter, replace the sniffer tip first.
If this does not cure the problem, the capillary in the sniffer line is clogged and the
complete sniffer line needs to be replaced.
10
Please recalibrate the Ecotec E3000 for most accurate display of leak rate.
8.7
Switching the capillary filter
At the top of the sniffer tip two different capillary filters can be mounted.
Fig. 8-12 Sinter filter
Fig. 8-13
Pos.
Description
Pos. Description
1
Taper Gasket (can be black or white)
3
Plastic capillary filter
2
Metal capillary filter
1
2
3
Содержание Ecotec E3000
Страница 2: ...0 2 kina22eIVZ fm Technical Handbook 1206 ...
Страница 13: ...General Information 1 3 kina22e chapter 1 fm technical handbook 1206 ...
Страница 30: ...1 20 General Information kina22e chapter 1 fm technical handbook 1206 ...
Страница 38: ...2 8 How the Instrument Works kina22e chapter 2 fm technical handbook 1206 ...
Страница 58: ...3 20 Operation of the Ecotec E3000 kina22e chapter 3 fm technical handbook 1206 ...
Страница 98: ...4 40 Equipment Settings kina22e chapter 4 fm technical handbook 1206 ...
Страница 106: ...6 8 Ecotec E3000 Messages kina22e chapter 6 fm technical handbook 1206 ...
Страница 133: ...Maintenance 8 18 kina22e chapter 8 fm technical handbook 1206 ...
Страница 143: ...Annex 9 10 kina22e chapter 9 fm technical handbook 1206 9 2 Declaration of Conformity ...