Ecotec E3000 Messages
6-3
ki
n
a
22
e c
h
ap
te
r 6
.fm
te
chn
ica
l ha
nd
boo
k
(1
20
6)
E25
Remove sniffer from CAL port
Sniffer inserted in calibration port during
start-up
or
sniffer needs to be removed during
internal calibration process
Remove sniffer from calibration port
of ECO-Check!
Light barrier of ECO-Check dirty
Blow out with fresh air and / or clean
with cotton swab!
W28
Real time clock reset! Please enter
date and time!
CPU-board has been replaced
Please enter date and time!
See Section
Battery on CPU-board faulty
Replace CPU-board*
E30
Sensitivity too low
Argon signal too low
Check if fresh air is provided to
sniffer line
Please recalibrate the Ecotec E3000
W31
Factor K1 out of range!
Interfering gases other than
cyclopentane or isopentane (e.g.
alcohols) detected
Recalibrate IGS! (see Section
W34
Flow has changed!
Flow has changed by more than 30%
since last calibration (warning will
disappear if change drops below 20%
again).
Recalibrate the
Ecotec E3000
(See Section
) or replace sniffer
line filters!
W35
Flow through capillary is too low
Actual flow is lower than lower flow limits
(In
LOW
FLOW
mode)
Filter in sniffer line blocked
Capillary blocked
Filter in main unit blocked
Lower flow limit setting too high
Replace sniffer tip filter
See Section
Replace filters of sniffer line and
recalibrate! (felt filter, capillary filter
and / or filter pad in SL3000XL
sniffer line and recalibrate) See
Section
or
Replace probe cable*
Replace internal filter*
Decrease lower flow limit
See Section
E36
Capillary broken (>2 s)
Actual flow is higher than upper flow
limits for more than 2 s
Capillary broken or leaky
Upper pressure limit setting too low
Replace sniffer line and recalibrate!
Or: Replace probe cable* and
recalibrate!
Increase upper flow limit (see
Section
Press OK to restart the Ecotec
E3000
E37
Capillary broken (>10 s)
Actual flow is higher than upper flow
limits for more than 10 s, emission will be
switched off to protect filaments
Capillary broken or leaky
Upper pressure limit setting too low
Replace sniffer line and recalibrate!
Or: Replace probe cable* and
recalibrate!
Increase upper flow limit (see
Section
Error no.
Plain text message
Possible reason
Recommendation
Содержание Ecotec E3000
Страница 2: ...0 2 kina22eIVZ fm Technical Handbook 1206 ...
Страница 13: ...General Information 1 3 kina22e chapter 1 fm technical handbook 1206 ...
Страница 30: ...1 20 General Information kina22e chapter 1 fm technical handbook 1206 ...
Страница 38: ...2 8 How the Instrument Works kina22e chapter 2 fm technical handbook 1206 ...
Страница 58: ...3 20 Operation of the Ecotec E3000 kina22e chapter 3 fm technical handbook 1206 ...
Страница 98: ...4 40 Equipment Settings kina22e chapter 4 fm technical handbook 1206 ...
Страница 106: ...6 8 Ecotec E3000 Messages kina22e chapter 6 fm technical handbook 1206 ...
Страница 133: ...Maintenance 8 18 kina22e chapter 8 fm technical handbook 1206 ...
Страница 143: ...Annex 9 10 kina22e chapter 9 fm technical handbook 1206 9 2 Declaration of Conformity ...