INFICON
Removing and Installing the Sensor | 15
074-708-P1A D-TEK Stratus OM
23 / 35
15 Removing and Installing the Sensor
D-TEK Stratus uses a cartridge style sensor that is quick and easy to remove and
replace in the field. In addition to the standard sensor, INFICON offers a CO
2
specific
sensor for use in refrigeration and air conditioning applications. See Replacement
eject button
sensor release
tab
To replace the sensor:
1
Press the eject button on the back of D-TEK Stratus and remove the battery
door.
2
Grasp the sensor release tab and gently pull it out.
3
Align the new sensor with the rails.
4
Gently push the sensor along the rails until it is fully inserted.
ð
Do not force the sensor. If it does not slide freely, check the alignment and try
again
5
Reinstall the battery door.