6 | Operation
INFICON
40 / 68
Contura® S400-Operating-instructions-jina91en1-09-(2001)
• Pac-Check (Catalog number 572-000).
The Pac-Check allows you to calibrate the leak rate.
• Calibration Kit (Catalog number 573-000).
The calibration kit lets you calibrate the leak rate and in addition the inner and
outer volumes.
6.10.3 Leak rate calibration
ü
Supervisor - rights
ü
You have a Pac-Check.
1
> Calibration
ð
The "Calibration" window opens in which you can calibrate "Leak rate", "Inner
volume" and "Outer volume".
2
To calibrate the leak rate using a calibration leak, select "Leak rate" in the
selection field and enter the leak rate of the calibration leak in mbar l/s in the
next field.
3
Press
and follow the instructions on the screen.
ð
You will now be prompted to perform a ZERO measurement and then to
measure the calibration leak.
ð
After calibration the determined calibration factor is displayed in the window
"Calibration".
6.10.4 Inner volume calibration
The internal volume is evaluated in the measuring modes “Small & Rigid" and "Large
& Rigid" for the large scale detection, see also "Determine the Measuring mode
ü
Supervisor - rights
ü
You have a blue specimen rod from the calibration kit (setting inner volume 20 ml).
1
> Calibration
ð
The "Calibration" window opens.
2
Select "Inner volume" in the selection field and then follow the instructions on
the screen.
ð
After the ZERO measurement, place the test specimen in the center of the
measuring chamber.
3
Press
.