INFICON
Operation | 6
UL6000-Fab-Operating-instructions-iina75en1-04-(2109)
87 / 146
6.10.1 Update the user interface software
Import software updates using a USB flash drive.
NOTICE
Loss of data due to disconnection
► Do not switch off the device and do not remove the USB flash drive while the
software is being updated.
ü
Supervisor rights
1
Copy the file into the main directory of a FAT32 formatted USB flash drive.
2
Connect the USB flash drive to a USB port on the device.
3
> Update > Update operating unit
ð
At the top in the window the active software version of the user interface is
shown.
If one or more versions of the software are on the USB flash drive the most
recent version is shown on the line below. If this is the same as the version
already installed the background is green, otherwise it is red.
4
In order to load the new software version, press on the button "Update".
ð
After completion there is an automatic restart of the operating unit.
6.10.2 Updating the software of the basic unit
Import software updates using a USB flash drive.
NOTICE
Loss of data due to disconnection
► Do not switch off the device and do not remove the USB flash drive while the
software is being updated.
ü
Supervisor rights
1
Copy the file into the main directory of a FAT32 formatted USB flash drive.
2
Connect the USB flash drive to the USB port on the device.
3
> Update > Update basic unit
ð
At the top in the window the active software version of the basic unit is
shown.
If one or more versions of the software are on the USB flash drive the most
recent version is shown on the line below. If this is the same as the version
already installed the background is green, otherwise it is red.
4
In order to load the new software version, press on the button "Update".
Содержание 550-520
Страница 2: ...INFICON GmbH Bonner Strasse 498 50968 Cologne Germany ...
Страница 119: ...INFICON Decommissioning 9 UL6000 Fab Operating instructions iina75en1 04 2109 119 146 ...
Страница 139: ...INFICON Appendix 11 UL6000 Fab Operating instructions iina75en1 04 2109 139 146 11 2 CE Declaration of Conformity ...
Страница 140: ...11 Appendix INFICON 140 146 UL6000 Fab Operating instructions iina75en1 04 2109 11 3 RoHS ...
Страница 143: ...INFICON Index UL6000 Fab Operating instructions iina75en1 04 2109 143 146 Z ZERO 77 Button 26 Factory setting 70 ...
Страница 144: ...Index INFICON 144 146 UL6000 Fab Operating instructions iina75en1 04 2109 ...
Страница 145: ......
Страница 146: ......