6.10 Switch off
To switch the device off, disconnect it from the power supply.
NOTICE
Material damage due to helium contamination
Helium remaining in the sensor shortens the service life of the sensor and increases
the warm-up phase.
► Do not switch off the device if the sensor is contaminated with helium. Wait until the
device is ready to measure again (measurement view is displayed).
6.11 Calling up information on the device and on the
measurement
All important information for operation can be displayed on the info menu.
► Select “Info” in the main menu.
Displaying/changing
settings
You can find all the current measurement settings under “View/change settings”. You
can select a setting using the right arrow key and switch to the corresponding setting
window by selecting “Change”.
Settings overview
Fig. 25: Settings overview
Leave the window by selecting “Cancel”. Changed settings are saved.
Internal data
You can display all of the device’s internal data by selecting “Internal data”. You can
switch between the 8 windows using the lower right and left keys .
Vacuum schematic
The image called up displays the device’s vacuum schematic and its valve
arrangement. In addition, sensor pressure, sensor current and an estimated helium
concentration in ppm are displayed.
Interfaces
You can display all information on the interfaces by selecting “Interfaces”. You can
select a setting using the right arrow key and switch to the corresponding setting
window by selecting “Change”.