5.2.2 Carrier gas measurement setup
Select the most helium-free environment possible for the measurement setup. For
reliable measurements with the T-Guard2, the helium content in the air must be less
than 10 ppm. By nature, air contains 5 ppm (0.0005%) helium.
1
2
3
4
5
6
7
8
9
Fig. 10: Measurement setup for carrier gas measurements
1
Carrier gas pump
6
Test object
2
Mass flow controller (optional)
7
Outside air inlet
3
Measurement line
8
Reference line
4
Measuring chamber
9
Fore pump
5
Fan
• Carrier gas pump
The carrier gas pump ensures carrier gas flow through the measurement
chamber. The following limits apply:
Expected leak rate
Carrier gas flow
1 × 10
-5
mbar l/s
<3000 sccm
1 × 10
-4
mbar l/s
<30000 sccm
1 × 10
-3
mbar l/s
<300000 sccm
• Mass flow controller (optional)
The mass flow controller controls the actual current mass flow. This allows for the
determination of a more accurate leak rate than an estimation of the mass flow.
• Measurement line