3000 Micro GC Operating Manual
2.3.5 Repeatability
Typical results for Relative Standard Deviation (RSD) at constant temperature and
pressure (for C
1
to C
6
components at % level) are shown below for peak areas.
Injector Type
Peak Area Repeatability
Variable volume . . . . . . . . . . . . . . . . <1% RSD
Fixed volume . . . . . . . . . . . . . . . . . . <0.2% RSD
Backflush (timed mode) . . . . . . . . . . <1% RSD
Backflush (fixed mode). . . . . . . . . . . <0.5% RSD
2.3.6 Carrier Gas Specifications
Connection to 3000 Micro GC . . . . . 1/8 in. Swagelok fittings
External source . . . . . . . . . . . . . . . . Helium
Hydrogen
Nitrogen
Argon
Carrier gas purity . . . . . . . . . . . . . . . 99.995 to 99.9995%
Input pressure . . . . . . . . . . . . . . . . . 552 to 566 kPa (80 to 82 PSI)
2.3.7 Portable 3000 Micro GC Carrier Gas Specifications
Onboard cylinder source . . . . . . . . . Helium
Nitrogen
Argon
Onboard cylinder capacity . . . . . . . . 12,410 kPa (1800 PSI)
Onboard cylinder usage . . . . . . . . . . 35-40 hours, application dependent
2.3.8 External Input/Output
RJ-45 based LAN
Power supply input connector
Remote start
2.3.9 Portable 3000 Micro GC Onboard Battery
The 3000 Micro GC onboard battery will operate the instrument up to two hours,
depending on the application.
Содержание 3000 Micro GC
Страница 1: ...O P E R A T I N G M A N U A L 3000 Micro GC Gas Analyzer PN 074 519 P1C Cover Page ...
Страница 2: ......
Страница 6: ......
Страница 8: ......
Страница 18: ...TOC 10 PN 074 519 P1C 3000 Micro GC Operating Manual This page is intentionally blank ...
Страница 26: ...1 8 PN 074 519 P1C 3000 Micro GC Operating Manual This page is intentionally blank ...
Страница 128: ...6 20 PN 074 519 P1C 3000 Micro GC Operating Manual This page is intentionally blank ...
Страница 268: ...11 20 PN 074 519 P1C 3000 Micro GC Operating Manual This page is intentionally blank ...