Table A.3 LEL correlation factors
a
for the sensor 17155304-U
Calibration gas
Sample gas
LEL
(% vol)
Propane
Acetone
2.5
3.28
Butane
1.9
0.97
Chloromethane
8.1
0.966
Cyclopentane
1.1
1.62
Dichloroethane
5.4
8.57
Ethane
3.0
1.01
Ethanol
3.5
1.65
Ethyl Acetate
2.0
1.69
Ethylene
2.7
3.43
Ethylene Oxide
3.0
0.845
Hexane
1.1
0.8
Isopropanol
2.0
1.43
Methane
5.0
3
Methanol
6.0
2.22
Methyl ethyl ketone
1.4
1.87
Pentane
1.4
0.89
Propylene
2.4
1.69
Toluene
1.1
1.18
Xylene
1.1
1.51
a
These factors only apply to gas concentrations expressed in % volume terms and up to 2.5%vol. These factors may vary from
sensor to sensor with tolerance of ± 25% deviation.
Note:
LEL correlation-factor accuracy may change without notice and is impacted by exposure to sensor inhibitors or poisons,
sensor aging, the gas-detection applications and environment, and other factors. Calibrate instruments using the intended target
gas when feasible and validate correlation factors as needed.