Table 4.1 Customer-supplied equipment and services
Item
Purpose
Restrictions
a
Mounting (optional)
Mounting surface and
mounting equipment items
Mounting the unit for
operation.
The mounting surface must support 10.9 kg (24 lb) static
weight and must meet any weight-bearing requirements
of the customer-supplied mounting equipment.
A strut-type kit or strut-type rails and fasteners for
mounting are suitable for a permanent installation. A unit
can also be surface mounted.
Wall Mount Kit from Industrial
Scientific
For use with surface mounting
or strut-type mounting
See the
section of this manual for instructions.
Note:
The kit includes a label that replicates important
information printed on the back of the RGX, including the
LAN MAC and WAN MAC values. Copy these two values
onto the kit label in the designated spaces. Affix the label
to an area of the unit that will be visible during operation.
Magnet Mount Kit from
Industrial Scientific
For use with metal surface
mounting
Do not
use with units that will draw power from a fixed DC
power source.
See the
section of this manual for instructions.
Power source, external (optional)
Fixed DC power source
Operational power suitable for
permanent installations.
9
–
30
VDC with a maximum current of 5.0A.
Hubs and conduits suitable for the application and hole
size 27.8 mm
(1.09 ″).
ISERTPS
b
from Industrial
Scientific
Operational power suitable for
extended run-time
installations.
Before using the ISERTPS, ensure your application
meets the product's certified uses and restrictions: read
and understand the ISERTPS
Product Manual (
part
number 17158248).
ERTPS
b
from Industrial
Scientific
Operational power suitable for
extended run-time
installations.
Before using the ERTPS, ensure it suits your application:
read and understand its
Product Manual
(part number
17158385).
Note:
Except for its IS cable, the ERTPS is
not
certified for use in any hazardous-classified areas.
Supplies (optional)
Torx screwdriver set with T20,
T25, and T30 bits
Removal or installation of
conduit-hub plugs (T20
screwdriver and 3/8 ″ wrench),
wall mounts (T30), and case
fasteners (T25)
Suitable for the installation location.
3/8 ″ open
-end wrench
Other (optional)
Ethernet cable (if needed, 90°
adapter or short adapter
cable)
Required only for units that
will use Ethernet during
operation.
Cat5 (or greater) Ethernet cable, terminating in RJ45
modular connectors, that supports 10/100 Ethernet. For
longer cables, 14−110 m [46–
360 '], use a solid
conductor shielded twisted pair cable. To prevent
damage to the Ethernet cable and the RGX, it is
recommended that the Ethernet cable have
either no
strain relief or flexible strain relief; if needed, a 90°
adapter or short adapter cable may be used.