IN USA IN2000-L2-LC Скачать руководство пользователя страница 1

 

Title:  Model IN2000-L2-LC Low Concentration Ozone Analyzer 
          Operating and Maintenance Instructions

 

Doc.#: 

610-0023-01 

Rev.: 

C

 

Proprietary:  The contents of this document are copyright protected.  
© 1998 by IN USA, INC.

 

Page 1 of

 30 

 

 

 

Date ECO# 

Rev 

Description 

Originator 

DEC-98  

1.0 FIRST 

RELEASE 

VJD 

Mar-99 

 

1.1 

Hi-Lo Alarm Capability 

VJD 

Sep-99 

 

1.2 

Flow rate for gas phase measurement 

VJD 

Oct-00 

172 

Release to Revision A 

CRH 

Nov-00 

189 

Updated Figure 4, Figure 7 

LZ 

OCT-26 

311 

Add. Addendum #1069 (pages 29,30) 

LZ 

 
 
 
 
 
 
 
 
 
 

TM

 

 
 
 
 
 
 

MODEL IN2000- L2-LC 

LOW CONCENTRATION 

OZONE ANALYZER 

 
 
 

Operating and Maintenance Instructions 

 
 

 

 

Содержание IN2000-L2-LC

Страница 1: ...USA INC Page 1 of 30 Date ECO Rev Description Originator DEC 98 1 0 FIRST RELEASE VJD Mar 99 1 1 Hi Lo Alarm Capability VJD Sep 99 1 2 Flow rate for gas phase measurement VJD Oct 00 172 A Release to R...

Страница 2: ...EV PRE AMP REV KEYBOARD REV SOFTWARE VERSION CALIBRATION DATE INITIALS OPERATING MANUAL VERSION TUBING TYPE TEFLON STAINLESS STEEL AC VOLTAGE SERVICE SPECIFIED VOLTAGE ___________________ CONFIGURATIO...

Страница 3: ...RAL DESCRIPTION 7 MECHANICAL INSTALLATION 7 ELECTRICAL INSTALLATION 13 Power Connections 13 Signal Connections Field Wiring 13 SAMPLE GAS CONNECTIONS 16 SETTING THE SAMPLE FLOW RATE 16 FRONT PANEL DES...

Страница 4: ...he contents of this document are copyright protected 1998 by IN USA INC Page 4 of 30 Alarm 1 and Alarm 2 22 Input Output 23 Operating Parameters 23 Time and Date 23 Programming Programmable Parameters...

Страница 5: ...tandard Traceable to the US NIST 1 Ozone Concentration Units Factory Set Gas Phase PPMV others upon special request Dissolved O3 mg L operated as part of a W 1 System Readout 2x20 character alpha nume...

Страница 6: ...bility or manufacturability Make sure that this manual is used with the original product it was shipped with Although every effort has been made to ensure accuracy of the information contained in this...

Страница 7: ...in the measurement cell MECHANICAL INSTALLATION The L2 LC can be housed for bench top 19 rack and wall mounting Figure 1 and Figure 2 provide mechanical dimensions for bench top and for 19 rack mounti...

Страница 8: ...ne Analyzer Operating and Maintenance Instructions Doc 610 0023 01 Rev C Proprietary The contents of this document are copyright protected 1998 by IN USA INC Page 8 of 30 Figure 1 Bench Top Dimensions...

Страница 9: ...alyzer Operating and Maintenance Instructions Doc 610 0023 01 Rev C Proprietary The contents of this document are copyright protected 1998 by IN USA INC Page 9 of 30 Figure 2 19 Rack Dimensions 19 00...

Страница 10: ...on Ozone Analyzer Operating and Maintenance Instructions Doc 610 0023 01 Rev C Proprietary The contents of this document are copyright protected 1998 by IN USA INC Page 10 of 30 Figure 3 Wall Mounted...

Страница 11: ...one Analyzer Operating and Maintenance Instructions Doc 610 0023 01 Rev C Proprietary The contents of this document are copyright protected 1998 by IN USA INC Page 11 of 30 Figure 3 Wall Mounted NEMA...

Страница 12: ...one Analyzer Operating and Maintenance Instructions Doc 610 0023 01 Rev C Proprietary The contents of this document are copyright protected 1998 by IN USA INC Page 12 of 30 Figure 3 Wall Mounted NEMA...

Страница 13: ...VAC 50 60 Hz power outlet The instrument requires up to a maximum of 70 Watts and is fused with two 75 Amp Slo Blo fuses found in the power receptacle located in the rear panel The NEMA 4X version ref...

Страница 14: ...y Contact Closed on Error Condition 3 8 Alarm 2 Relay Contact Open on Alarm Condition 4 9 Alarm 2 Relay Contact COMMON 4 10 Alarm 2 Relay Contact Closed on Alarm Condition 4 11 Alarm 1 Relay Contact O...

Страница 15: ...a Low impedance short circuit protected analog output intended primarily for data acquisition recording and monitoring 2 4 20 mADC is a non isolated current loop Compliance is 750 Ohm and the loop is...

Страница 16: ...t only the gas sample is taken into the tubing and not any liquid or particulate matter It is highly recommended the sample line be filtered using one of the filters provided by IN USA In the NEMA 4X...

Страница 17: ...Page 17 of 30 FRONT PANEL DESCRIPTION The front panel of the Bench Top and 19 units is illustrated in Figure 5 The front panel contains the digital readout keyboard and dual flow meters Figure 5 Fron...

Страница 18: ...isplayed information is updated every 2 seconds Keyboard Description There are 5 active keys located in the front panel of the L2 LC analyzer 1 SCROLL identified as a circular arrow 2 DIGIT INCREMENT...

Страница 19: ...e ports in the NEMA version of the unit Note that there are 2 two electrical entry ports identified as FIELD WIRING and MAIN POWER intended for allowing electrical conductors to penetrate the enclosur...

Страница 20: ...e Teflon element filter supplied by IN USA to remove particulate matter from the gas sample Turning the Unit On and Warming Up Turn the instrument on by actuating the Power switch located in the rear...

Страница 21: ...ering and Program change several parameters The instrument is normally in Review mode and may be put into Program mode if a parameter needs to be changed See the next section on programming programmab...

Страница 22: ...Latched alarm is accomplished by pressing the E key located on the front panel Of course pressing the E key will clear a Latched alarm only if the original alarm condition is no longer present Input O...

Страница 23: ...log output The L2 LC must be in Program mode to access simulation Operating Parameters Time and Date This is a factory installed option Refer to the Manual Addendum of this Option for details on Time...

Страница 24: ...h Low Speaker Enabled Disabled 2 NUMERICAL The value of each digit in a numerical parameter can be changed For instance Alarm 1 Threshold xxx x digits Time xx xx xx To change the value of a Toggle par...

Страница 25: ...end The main display consists of the ozone temperature pressure and Review System Parameters screens These can be viewed separately by using the D key Figure 9 Menu Structure Scroll E Review System Pa...

Страница 26: ...rigger the Instrument Error Relay Once the causing event is not longer present this conditions require user intervention to extinguish Via a front panel keyboard E Key or through the RS 232 interface...

Страница 27: ...ES TO HANDLE THE UV LAMP To replace the UV lamp perform the following steps 1 Remove the cover from the unit 2 Disconnect the UV lamp from its power supply printed circuit board 3 Loosen the set screw...

Страница 28: ...lves are NIST traceable A certificate of calibration is issued with each unit Further details on this procedure are available from IN USA Inc Many end users opt to send instruments to our facility on...

Страница 29: ...232 port Refer to the Operation Manual for Document 610 0023 01 for RS 232 Port location and Pin out Analog Output High Setting Analog Output Low Setting Programming of Baud Rate The data structure i...

Страница 30: ...nding another F Command will result in stopping the dump of the message illustrated above To calculate the concentration c in ppmV from the RS232 dump simply compute c ppmV Y X Z ZZZ For instance Rang...