In-situ RDO PRO-X Скачать руководство пользователя страница 25

25

Calibration Slope and Offset

These values represent the slope and offset that will be applied to the raw concentration reading from the sensor to generate
the final values reported by the sensor parameters. These registers may be written independently of the normal internal
calibration procedure.

Entering Calibration Registers

The sensor is calibrated using the following procedure.

1. Optional: Read the Sensor Data Cache Timeout register 9463 and store the value.

2. Write the Sensor Data Cache Timeout register 9463 to a value less than your intended sample rate and greater

than 1000 milliseconds. This will ensure that you get new sensor readings during the stabilization process.

3. Optional: Read the temperature units register 0049 and saturation units register 0041 and store their values.

4. Write the temperature units register 0049 to its default value (1) and write the saturation units register 0041 to its

default value (117).

5. Write the Calibration Mode On command (0xE000) to the sensor command register 9305.

6. Update the live salinity and barometric pressure registers if necessary.

7. Prompt the user to place the probe in a 100% saturation environment.

8. Read the oxygen concentration and temperature parameters. When these values have reached equilibrium, record

them in their respective 100% saturation calibration registers. Write the current live salinity and barometric
pressure readings to their respective calibration registers.

9. Place the sensor in a 0% saturation environment. When these registers have reached equilibrium, record them in

their respective 0% saturation calibration registers. If a zero calibration is not to be performed, these registers can
be set to zero or left at their previous values.

10. Write the Calibration Update command (0xE001) to the sensor command register. The sensor will calculate a new

slope and offset, write the current time to the last user calibration time register, and set the next user calibration
time register to zero (disabled). If the concentrations at 100% and 0% saturation are equal, the probe will return an
exception response with code 0x97 (invalid calibration) and not attempt to compute a new slope and offset due to
possible division by zero. If the slope does not calculate between 0.85 and 1.20 inclusive, or if the offset does not
calculate between -0.2 and +0.2 inclusive, then the probe will return an exception response with code 0x97 (invalid
calibration).The slope and offset will be available for read but will not be committed to flash.

11. Optional: Read the last user calibration time register, add the next calibration interval, and write the result to the

next user calibration time register.

12. Write the Calibration Mode Off command (0xE002) to the sensor command register to place the sensor in normal

operation. If the calibration mode is turned off without a calibration update command, or the calibration command
returned an exception, the previous calibration shall be restored.

13. Optional: If you saved the temperature and saturation parameter units at the start of the process, write the original

values back.

14. Optional: If you saved the Sensor Data Cache Timeout register 9463 at the start of the process, write the original

value back.

Содержание RDO PRO-X

Страница 1: ...itu In Situ logo Baro Merge BaroTROLL HERMIT HydroVu iSitu Pocket Situ RDO RuggedCable RuggedReader SmarTROLL TROLL VuSitu and Win Situ are trademarks or registered trademarks of In Situ Inc 2008 2015...

Страница 2: ...is a registered trademark of Intel Tefzel and Delrin are registered trademarks of E I DuPont de Nemours and Company Viton is a registered trademark of DuPont Dow Elastomers Kellems is a registered tra...

Страница 3: ...tain Desiccant 13 6 Controller Requirements and Connections 14 Wiring Overview 14 Analog 4 20 mA 3 wire 15 SDI 12 3 wire 16 Modbus Master with Built in RS485 17 Modbus Master with Built in RS232 Conve...

Страница 4: ...bration Slope and Offset 25 Entering Calibration Registers 25 Calibration Calculations 26 10 Service 27 Sensor Health Table 27 11 Appendix A Communication Device 28 Install and Open the Software 28 Co...

Страница 5: ...s Modbus RS485 4 20 mA SDI 12 Max power consumption 50 mA at 12 VDC Measure current 6 mA typical at 24 VDC Idle current no measurement or communication 160 A typical at 24 VDC Maximum cable length Up...

Страница 6: ...technology for DO measurement System Description The RDO PRO X system consists of the following items l Black probe body with removable nose cone l Optical RDO X Sensor Cap Classic Cap or Fast Cap l T...

Страница 7: ...he Probe 1 Remove the probe from the box and other packaging materials 2 Unscrew the nose cone from the probe and remove the red protective dust cap from the sensor Save the dust cap for later use 1 D...

Страница 8: ...midity inside the cap Keep the cap in its sealed package until you are ready to install it Install promptly Ensure that O ring grooves are dry and that the O ring is not rolled or pinched inside the c...

Страница 9: ...with approximately 10 mL water 3 Gently dry the instrument and sensing material with a paper towel making sure there is no water or debris on the instrument or on the sensing surface 4 Place the inst...

Страница 10: ...n 1 Fill line 2 Temperature sensor 2 Place the instrument into the solution Leave at least 13 mm 0 5 in between the surface of the sensing material and the bottom of the chamber 3 Ensure that the temp...

Страница 11: ...RDO PRO X Probe is internally threaded 1 11 NPT and can be attached to a male threaded pipe When deployed make sure that the nose cone and thermistor are completely submerged 1 10 m cable attached 2 T...

Страница 12: ...rial Clean the Probe With the RDO cap installed on probe gently scrub the probe with a soft bristled brush or nylon dish scrubber Use Alconox to remove grease or other matter Soak in vinegar and DI wa...

Страница 13: ...Desiccant installed in a controller or transceiver protects probe electronics from condensation A desiccant pack changes color from blue to pink as it becomes saturated with moisture It is extremely i...

Страница 14: ...85 Modbus l RS232 to Modbus Wiring Overview Refer to diagrams on the following pages Trim and insulate unused wires The shielded wire should be connected to a chassis ground or earth ground The inside...

Страница 15: ...3 wire Signal Color Ground Return Black External Power 12 36 VDC Red 4 20 mA Brown Cable length must not exceed 1219 m 4000 ft Analog signal must be enabled in Win Situ 5 Software or the Comm Kit Soft...

Страница 16: ...16 SDI 12 3 wire Signal Color Ground Return Black External Power 9 6 16 VDC Red RS485 Green RS485 Blue SDI 12 White Cable length must not exceed 60 96 m 200 ft...

Страница 17: ...17 Modbus Master with Built in RS485 Signal Color Ground Return Black External Power 12 36 VDC Red RS485 Green RS485 Blue Cable length must not exceed 1219 m 4000 ft...

Страница 18: ...Required Signal Color Ground Return Black External Power 12 VDC voltage limited by converter Red RS485 Green RS485 Blue Cable between converter and master must not exceed 60 96 m 20 ft Cable between m...

Страница 19: ...ut can be continuous in Modbus or SDI 12 mode as long as Modbus device register 9507 is set to 1 Communications The device automatically switches between Modbus and SDI 12 modes depending on which of...

Страница 20: ...Status Registers Register Size Mode Access Level R W Data Type Description 0005 3 R1 time Cap start date time 0 no cap 0008 3 R1 time Cap end of usable life date time 0 no cap Device Specific Register...

Страница 21: ...C default 2 F 0050 1 R1 ushort Data quality ID 0051 2 R1 W3 float Offline sentinel value default 0 0 0053 1 R1 16 bits Available units 0x00030 3 Dissolved Oxygen Saturation 0054 2 R1 float Measured v...

Страница 22: ...22 Dissolved Oxygen Equations Dissolved Oxygen Concentration DO concentration is internally calculated in mg L Conversion to other units is as follows...

Страница 23: ...23 Dissolved Oxygen Saturation...

Страница 24: ...etric pressure is used in the calculation of percent saturation and to determine the theoretical saturation point during calibration Values must be written in millibars in the range 506 625 to 1114 67...

Страница 25: ...se registers have reached equilibrium record them in their respective 0 saturation calibration registers If a zero calibration is not to be performed these registers can be set to zero or left at thei...

Страница 26: ...26 Calibration Calculations...

Страница 27: ...Health Table The following error messages appear in the software and the data file when a sensor issue has been detected Abbreviation Data Quality ID Text Description None 0 None Normal Data Quality U...

Страница 28: ...ce The Communication Device connects a stripped and tinned probe to a computer via USB connection 1 Disconnect the instrument from the PLC 2 The communication device includes an electrical connection...

Страница 29: ...29 Declaration of Similarity...

Отзывы: