Instruction Sheet

Care and Maintenance of Aqua TROLL

®

and Level TROLL

®

Instruments

User Maintenance

Overview

It is important for users to perform scheduled
maintenance on their instruments to sustain the
accuracy and longevity of the probes and cables. The
frequency of this maintenance depends on the
characteristics of the deployment site, including
humidity levels and the degree of fouling.

Users should be aware of the conditions at their
deployment sites and develop appropriate maintenance
schedules to replace desiccant, clean the instruments,
and send in the instruments for factory calibration.
Users should check instruments often during the first
portion of the deployment to determine the frequency of
maintenance. General maintenance should be
performed as often as possible. Factory service should
be performed at least every 12 months.

Preventing Condensation

Desiccant

Desiccant protects electronics from condensation,
which can cause irreparable damage and loss of data.
Indicating desiccant changes from blue to pink as it
becomes saturated with moisture.

Desiccant stages (from left)
New, nearly expired (replace now), expired

It is

extremely important

to use a properly-sized

desiccant for your deployment, and to

change

desiccant often

. Desiccant should be changed before

the entire volume has turned pink, and you should use
enough desiccant to effectively keep cables,
instruments and electrical boards dry until your next
scheduled maintenance. Desiccant life span is
dependent on site conditions.

Desiccant Pack Options

Small Desiccant

The Small Desiccant is a disposable cap that ships with
In-Situ Inc. products. The small desiccant is meant to
protect the instrument and cable

only during

shipping, and should not be used for
deployments

.

Large Desiccant (PN 0051810 and 0053550)

The Large Desiccant is a replaceable desiccant pack
designed for low-humidity environments or
deployments where maintenance occurs frequently.
The large desiccant is available with a plastic or a
titanium twist-lock connector.

Information subject to change without notice. In-Situ, In-Situ logo, Baro Merge, BaroTROLL, HERMIT, Pocket-Situ, RDO, RuggedCable,
RuggedReader, TROLL, and Win-Situ are trademarks or registered trademarks of In-Situ Inc. © 2012. All rights reserved.

08/2012 | Rev.001 |

Содержание Aqua TROLL

Страница 1: ... it becomes saturated with moisture Desiccant stages from left New nearly expired replace now expired It is extremely important to use a properly sized desiccant for your deployment and to change desiccant often Desiccant should be changed before the entire volume has turned pink and you should use enough desiccant to effectively keep cables instruments and electrical boards dry until your next sc...

Страница 2: ...sing the Desiccant Refill Kit 1 Remove the black nylon vent cap from the top of the desiccant 2 Pour out and discard the used desiccant Check the glass wool in the bottom of the container and replace if necessary The glass wool prevents the desiccant beads from falling out the cable end of the desiccant pack If the wool glass does this effectively there is no need to replace it when refilling the ...

Страница 3: ...ch as a dilute solution of dish detergent Be careful not to damage the plastic material of the conductivity cell Instruments are shipped with polyurethane foam swabs for this purpose You can also use a thin cotton pipe cleaner and clean the cell with a gentle back and forth flossing motion Process 2 Light scrubbing with a foam swab and an aggressive soap such as Alconox detergent can be used for s...

Страница 4: ...equent fast sampling or during long term deployments When this power source is connected the TROLL instrument will use the external battery source first and switch to the internal batteries when external battery power is depleted Battery life depends on sampling rate This battery pack allows for 1 5 V UM 3 or size AA batteries 8 that are replaced by the user TROLL Replaceable Battery Pack Estimate...

Страница 5: ... 12 months or when the unit data appears to drift significantly Factory calibration includes a thorough cleaning full functionality check and adjustments of the pressure sensor temperature sensor and conductivity sensor when applicable over the entire calibrated temperature range To obtain a factory calibration fill out and return the online RMA Return Materials Authorization form located at www i...

Страница 6: ...s in your Operator s Manual or related In Situ Inc documentation may cause damage to your instrument your warranty to become void or both Please follow all instructions and warnings to avoid damage to your product The presence of the Waste Electrical and Electronic Equipment WEEE marking on the product indicates that the device is not to be disposed via the municipal waste collection system of any...

Отзывы: