30
Salt may accumulate around the reference junctions of the ammonium, chloride, nitrate and
pH sensors. Rinse with deionized water to remove any buildup.
Before using the
ISE sensors
for the first time, replace the reference filling solution. Condition
the sensors by soaking in calibration standard for 4-24 hours prior to deployment. This step is
not necessary for the pH sensor.
The wiper can severely reduce the life of the RDO Fast Cap. Wear will vary by application. Verify
sensor performance prior to use and replace the Fast Cap if damaged.
Handling pH and Ion-Selective Electrode Sensors (ISEs)
Using the RDO Sensor and RDO Fast Cap
Unscrew reference junction.
Replace reference filling solution.
1
2
Potential salt buildup. Rinse with deionized
water if necessary.
Inspect foil prior to each use. If
damaged, replace cap.
RDO Fast Cap
Wiper
1-970-498-1500
www.in-situ.com