NanoPhotometer

® 

P-Class 

User Manual 

P 300 / P 330 / P 360 

 

Version 2.1 

 
 
 
 
 

P 360 

P 300 

P 330 

Содержание NanoPhotometer

Страница 1: ...telephone 49 89 726 3718 0 Fax 49 89 726 3718 54 Email info implen de www implen de NanoPhotometer P Class User Manual P 300 P 330 P 360 Version 2 1 P 360 P 300 P 330...

Страница 2: ...e 1995 5 EC Radio and Telecommunications Terminal Equipment Directive instruments fitted with Bluetooth accessory only Standards to which conformity is declared where relevant are as follows EN 61010...

Страница 3: ...in Determination 19 4 2 1 General Information 19 4 2 2 Protein UV Method 20 4 2 3 Protein UV Dye Method 22 7 3 4 BCA Assay 24 7 3 5 Bradford Assay 27 7 3 6 Lowry Assay 30 7 3 7 Biuret Assay 33 4 3 Bac...

Страница 4: ...shooting 64 9 ACCESSORIES 64 10 SPECIFICATION AND WARRANTY 65 11 SOLVENT COMPATIBILITY 66 12 APPENDIX 67 12 1 Nucleic acid quantification 67 12 2 Nucleic acid fluorescent dye incorporation 67 12 3 Pro...

Страница 5: ...has been established 2 3 hours Maximum relative humidity of 80 up to 31 C decreasing linearly to 50 at 40 C The instrument must be placed on a stable level bench or table that can take its weight 4 5...

Страница 6: ...or in native NanoPhotometer P Class Software format for later access see also NanoPhotometer P Class PVC Installation and User Manual Alternatively results may be saved on a SD Memory Card P300 only o...

Страница 7: ...levant section of the NanoPhotometer P Class User Manual Alphanumeric keys Use these to enter parameters and to write text descriptions where appropriate or required Use repeated key presses to cycle...

Страница 8: ...here appropriate or required Use repeated key presses to cycle through lower case number and upper case Leave for 1 second before entering next character Use C button to backspace and 1 to enter a spa...

Страница 9: ...art from 8 Save the parameters as a method 9 Open printer settings possibility to change the printer settings within the method as described in 7 3 Output Options Printer Exit Options by pressing Esca...

Страница 10: ...Photometer P Class Submicroliter Cell into the cell holder with the cell windows facing the light beam We recommend facing the Implen logo to the front The light beam is directed from RIGHT to LEFT as...

Страница 11: ...pplications and Cuvette Applications The procedure is as follows Exemplary Parameter Screen Parameter Screen Step 1 Press 1 to select NanoVolume Applications folder Step 2 Press 1 to select Nucleic Ac...

Страница 12: ...id 100 Lid 100 Concentration too low Do you want to change the lid factor Please change to lid 50 and press sample automatic change of lid factor lid 100 to lid 50 in the software for calculation No c...

Страница 13: ...e 10 mm pathlength Dilution factor and cell pathlength can be entered Nucleic Acid Purity Checks Nucleic acids extracted from cells are accompanied by protein and extensive purification is required to...

Страница 14: ...cent dye incorporation To determine the dye incorporation rate the absorbance reading at the wavelength reported for maximum absorbance of the fluorescence dye is used The corresponding extinction coe...

Страница 15: ...ng the left and right arrows Options g ml ng l g l Step 8 Enter the Factor using the keypad numbers Default value is 50 for dsDNA 37 for ssDNA and 40 for RNA range is 0 01 to 9 999 Step 9 Press OK to...

Страница 16: ...ction table denoting the ratios of the 4 bases according to the oligo sequence Enter the proportions of bases present using the keypad numbers and up and down arrows to move between boxes Default is 1...

Страница 17: ...to select one of the dye incorporation methods Step 4 Using the NanoVolume Applications select the Lid Factor as described in the Average Detection Range Sheet and under 3 2 Step 5 Select Dilution Fac...

Страница 18: ...f the absorbance value of the sample is not in the linear range a Warning message will pop up and Instruction will be displayed in the top left corner of the result screen Please refer to 3 2 Software...

Страница 19: ...cations and Cuvette Applications To determine the degree of labelling the absorbance reading at the wavelength reported for maximum absorbance of the fluorescence dye is used The corresponding extinct...

Страница 20: ...e keypad numbers Range 0 01 to 9 999 Press OK to calculate the dilution factor and return to the Parameters screen OR press Cancel to cancel the selections and return to the Parameters screen Step 5 S...

Страница 21: ...alue of the sample is not in the linear range a Warning message will pop up and Instruction will be displayed in the top left corner of the result screen Please refer to 3 2 Software instructions impo...

Страница 22: ...ein BSA default Serum Albumin mouse Serum Albumin human IgG Lysozyme Custom or OD 1 Step 8 If using Custom Protein there are two possibilities to enter the correct factors see also page 18 protein UV...

Страница 23: ...ulated Step 16 If the absorbance value of the sample is not in the linear range a Warning message will pop up and Instruction will be displayed in the top left corner of the result screen Please refer...

Страница 24: ...significant figures regardless of how many decimal points are selected so 98768 2 will display as 98768 even with 1 decimal point selected Press OK to store the chosen parameters OR Cancel Step 7 Pres...

Страница 25: ...measured press OK to accept the calibration and go to the Results screen see below OR press Back to cancel selections and return to the Standards screen Calibration Screen replicates on Step 17 This s...

Страница 26: ...id unintended escaping the application NanoPhotometer P 300 Options select using key pad numbers NanoPhotometer P 330 P360 Menu select using key pad numbers Options select using key pad numbers 1 Retu...

Страница 27: ...to 5 significant figures regardless of how many decimal points are selected so 98768 2 will display as 98768 even with 1 decimal point selected Press OK to store the chosen parameters OR Cancel Step...

Страница 28: ...ured press OK to accept the calibration and go to the Results screen see below OR press Back to cancel selections and return to the Standards screen Calibration Screen replicates on Step 17 This shows...

Страница 29: ...oid unintended escaping the application NanoPhotometer P 300 Options select using key pad numbers NanoPhotometer P 330 P360 Menu select using key pad numbers Options select using key pad numbers 1 Ret...

Страница 30: ...5 significant figures regardless of how many decimal points are selected so 98768 2 will display as 98768 even with 1 decimal point selected Press OK to store the chosen parameters OR Cancel Step 7 P...

Страница 31: ...ured press OK to accept the calibration and go to the Results screen see below OR press Back to cancel selections and return to the Standards screen Calibration Screen replicates on Step 17 This shows...

Страница 32: ...unintended escaping the application NanoPhotometer P 300 Options select using key pad numbers NanoPhotometer P 330 P360 Menu select using key pad numbers Options select using key pad numbers 1 Return...

Страница 33: ...5 significant figures regardless of how many decimal points are selected so 98768 2 will display as 98768 even with 1 decimal point selected Press OK to store the chosen parameters OR Cancel Step 7 P...

Страница 34: ...ured press OK to accept the calibration and go to the Results screen see below OR press Back to cancel selections and return to the Standards screen Calibration Screen replicates on Step 17 This shows...

Страница 35: ...oid unintended escaping the application NanoPhotometer P 300 Options select using key pad numbers NanoPhotometer P 330 P360 Menu select using key pad numbers Options select using key pad numbers 1 Ret...

Страница 36: ...t geometry of this slit and the monochromator optics Different spectrophotometer types therefore give different responses for the same turbid sample to compare results they must be normalized using ca...

Страница 37: ...ght arrows Options are 1 000 or 1 000 000 Factor and Multiplier define the conversion of the measured OD to the number of cells per millilitre e g 1 OD 600 5 x 108 cells ml Step 8 Press OK to enter th...

Страница 38: ...ed on a user programmed curve 6 Absorbance or T transmission at up to 5 user defined wavelengths 7 Ratio of absorbance values at two user specified wavelengths Menu Options Within each function the us...

Страница 39: ...ep 5 To enter the results screen with the selected parameters press OK OR cancel the selections and return to the Functions folder by pressing Cancel Results Screen Step 6 Insert the reference sample...

Страница 40: ...d enter name 9 Open printer settings possibility to change the printer settings within the method as described in 7 3 Output Options Printer Exit Options by pressing Escape or wait Menu select using k...

Страница 41: ...utton to delete the last digit entered Step 6 if Standard is selected Enter the concentration using keypad numbers Range 0 01 to 9 999 Use the C button to delete the last digit entered Step 7 Units Th...

Страница 42: ...pad numbers 1 Return to parameters screen 2 Print result via selected method 3 Toggles on off displaying a graph of wavescan 20 nm from selected wavelength 4 Return to Run Standard screen 7 Sample nu...

Страница 43: ...ncel Measurement Screen Step 7 Insert the reference sample Press Blank key This will be used for all subsequent samples until changed Step 8 Insert sample and press Sample Step 9 Repeat for all sample...

Страница 44: ...settings within the method as described in 7 3 Output Options Printer Exit Menu Options by pressing Escape or wait Peak Detection Shortcut button 4 Auto Detect Peaks Turns on and off the automatic pe...

Страница 45: ...aph Scale Shortcut button 6 This enables the user to set up a defined graph by defining the limits in either or both of the x and y axes Zoom mode This sets up the operation of the Zoom keys up and do...

Страница 46: ...minutes Step 5 Duration Enter the time in minutes over which measurements are taken This can be a maximum of 60 minutes Step 6 Interval Enter the interval time in seconds between measurements using th...

Страница 47: ...Query needs confirmation to avoid unintended escaping the application Menu select using key pad numbers 1 Return to parameter screen 2 Transfer the results via selected Output Options Printer setting...

Страница 48: ...e Menu Options key and then use the left right arrows g ml g l pmol l mg ml mmol l mol l g l mg l g l U l ppm ppb conc or none These units can also be edited once OK is pressed This screen also allows...

Страница 49: ...results and the fitted curve as the measurements are input Step 16 Use the up and down arrows to select a standard to be repeated if a poor reading has been obtained Use C to clear the previous readi...

Страница 50: ...ll subsequent samples until changed Step 28 Insert the sample and press Sample The concentration of the sample is taken and displayed Step 29 Repeat for all samples Step 30 Press Menu Options to displ...

Страница 51: ...bove and repeat for the number of wavelengths selected up to 5 Step 7 Press OK to enter the results screen OR press Cancel to return to the Functions folder Results Screen Step 8 Insert the reference...

Страница 52: ...tore in User Methods 1 9 press the down arrow and enter name 9 Open printer settings possibility to change the printer settings within the method as described in 7 3 Output Options Printer Exit Option...

Страница 53: ...ction is On Enter the third Wavelength from which the background correction will be obtained Step 7 Press Next to enter the next screen OR press Cancel to return to the Functions folder Step 8 Dilutio...

Страница 54: ...ions which are described below Step 18 Press Escape and confirm with Yes to return to the Functions folder Query needs confirmation to avoid unintended escaping the application Menu select using key p...

Страница 55: ...ws 4 Press OK to lock the method OR Cancel to return to the User Methods folder Unlock Method 1 Press 3 to select Unlock Folder 2 Select the method to be unlocked using the left and right arrows 3 Ent...

Страница 56: ...thod 3 Press OK to delete the method OR Cancel to return to User Methods folder 7 UTILITIES Survey of the available Utilities Key pad number Description 1 Set correct time and date 2 Select preferred...

Страница 57: ...s folder OR press Cancel to return to the Utilities folder without storing the settings 7 3 Output Options Printer 7 3 1 NanoPhotometer P 300 The NanoPhotometer P 300 offers a selection of data output...

Страница 58: ...ease proceed to Folder 5 Utilities press 3 Output Options to set the favoured reporting settings Data output to Built in printer Printing on the built in printer is possible independent from the Outpu...

Страница 59: ...urn to the Utilities folder without storing the settings If Computer USB is selected all measured data are sent automatically to the PC The file is closed and saved once the method is left by pressing...

Страница 60: ...urs at night or off Step 5 Select Baseline Compensation to improve value stability and to overcome background effects Step 6 Press OK to store the settings and return to the Utilities folder OR press...

Страница 61: ...ing GLP GMP calibration certification using filters traceable to international standards during production and quality control certified engineers and calibrated test equipment approved to ISO 9001 st...

Страница 62: ...isinfection Desinfection or cleaning only by wiping no spraying Use no solvents or aggressive chemicals Approved disinfectant solutions Apesin disinfection spray Tana Chemie GmbH Incidin Liquid and In...

Страница 63: ...n No light on Reference channel Submicroliter cell or cuvette in the cell holder by switching on the instrument Cell holder has been removed from the instrument and placed back in the wrong position R...

Страница 64: ...of paper normal sound no error message Thermal Printer paper was probably put the wrong way check correct position of the printer paper Thermal printer automatic roller stops Please contact the Implen...

Страница 65: ...n factors 5 10 50 100 and 250 Other technical data Vortexer 2 800 rpm tube size up to 2 0 ml Cuvette storage capacity for eight 10 mm cells Photometric mode Abs T concentration scan ratio multi wavele...

Страница 66: ...nzene butanol carbon tetrachloride chloroform ethanol ether HEPES hexane isopropanol MES methanol methylenchlorid MOPS phenol up to 1 n propanol toluene buffers containing phosphate or chloride salts...

Страница 67: ...on on and off Calculation of the fluorescence nucleic acid concentration c nuc Abs 260 Abs 320 factor nuc lid factor dilution factor With dye correction c nuc Abs 260 Abs 320 CF dye Abs max dye Abs 32...

Страница 68: ...cm 1 The following dye types and parameters are pre programmed in the NanoPhotometer P Class Dye Type Absorption maximum Dye nm Dye dependent extinction coefficient dye Dye dependent correction factor...

Страница 69: ...incorporation To determine the protein concentration and the dye concentration after labelling a modification of the Lambert Beer equation is used Background correction is always calculated possibili...

Страница 70: ...ependent molar extinction coefficient M 1 cm 1 The following dye types and parameters are pre programmed in the NanoPhotometer P Class Dye Type Absorption maximum Dyes nm Dye dependent extinction coef...

Отзывы: