NanoPhotometer
®
N120/
NP80/N60/N50/C40
User Manual Version 4.3.1
55
It is also possible to use cuvettes with 5 mm, 2 mm, 1 mm or 0.5 mm path lengths, but
there may be an adapter necessary. Please ask your cuvette supplier for a suitable
adapter.
The minimum volume for accurate measurements depends on the cuvette type used; it
is necessary that the light passes through the sample for accurate measurements.
Center height is 8.5 mm.
Note
: The cuvette holder is not removable. Do not pour any cleaning solution into the
cuvette holder as larger amounts of liquids can get into the instrument and cause damage.
S
OLVENT
C
OMPATIBILITY
(N50/N60/N120/NP80/C40
WITH
SUBMICROLITER CELL
)
Most solvents typically used in life science laboratories are compatible with the
NanoPhotometer
®
NanoVolume sample surfaces. The following solvents are compatible for
use with the NanoPhotometer
®
models N50, N60, N120 and NP80 at room temperature:
Acetone (
≤
5%)
Acetonitrile
Benzene
Butanol
Carbon tetrachloride
Chloroform
Ethanol
Ether
HEPES
Hexane
Isopropanol
MES
Methanol
Methylene chloride
MOPS
Phenol (
≤1%)
N-propanol
Toluene
Phosphate containing buffers
PBS (pH 4-10)
Citrate
Borate
Chloride salts
Acids
>
pH 2
Bases
<
pH 10
Note
: Highly concentrated acids and bases are not recommended. It is recommended to
wipe the sample surface with a lint-free laboratory wipe immediately upon completion of
each measurement. For more information about compatibility of specific solvents not listed
above, please contact the Implen support team (
) to check the
compatibility.